Quart Healthcare Topaz 8 Plus User Manual Download Page 11

10 

Pressing both the Soft and Firm buttons together will activate Auto-Detection 

and Auto-Detection will automatically set the appropriate pressure output for 
the patient. When activated, the Auto-Detection light will flash to indicate the 
Control Unit is detecting the comfort level for the individual. Once the Control 
Unit has completed detecting, the Auto-Detection light will stop flashing and 
remain ON. The mattress pressure can be manually adjusted by pressing Soft or 
Firm button if the individual wishes to change the comfort level. 

 

 

5.2

 

OPERATION GUIDE 

5.2.1

 

General Operation: 

NOTE: 

The power switch is located on the side of the Pump. 

 

Press

  to turn on the unit, all LED indicators on the control panel will light 

up accompanied with a beep for 2 seconds (check for indicator failure if any), 

and the Standby indicator on the control panel will light up. If the Control Unit 

was previously shut off in operate mode then the Control Unit will enter 

operate mode directly. 

P.S. To test the alarm battery, press 

  to turn off the power and the power 

failure alarm should be triggered. Refer to 

5.2.3 Audiovisual Alarm

 if the 

alarm is not triggered. 

 

Press the Operate button 

  and the system will begin to inflate and the 

"Auto-Firm" indicator will be flashing. 

 

The mattress should be fully inflated within 60 minutes, and automatically 
enter the previous operating mode, otherwise the low pressure alarm with 
warning code “

” will be triggered. 

  Note: Do not proceed to other settings before inflation is completed.   

 

After initial inflation is completed. Press Auto-Firm button 

  to transfer 

person onto the mattress. The mattress will turn into a static condition in 
around 5 minutes. Move the patient onto the mattress and press Auto-Firm 
button 

  again to cancel Auto-Firm mode and select the appropriate mode.   

 

According to the weight of the patient, adjust the pressure setting to the most 
suitable level without “bottoming-out”. User can determine an appropriate 
pressure by adjusting the Comfort Level. Please consult with your physician for 
a proper setting. 

Summary of Contents for Topaz 8 Plus

Page 1: ...Alternating Micro Low Air Loss Pressure Relief System User Manual Manufactured by Air Kinetic Technologies Corp Distributed by Quart Healthcare Inc 36 Northline Road Unit 9 Toronto ON M4B 3E2 www quarthealthcare com ...

Page 2: ...Please do not power off or unplug the Control Unit while in operation 2 Always use the same voltage as stated on the label Do not use other power cords on the Control Unit 3 Equipment is not suitable to use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide 4 Keep away from sharp objects 5 Close supervision is necessary when this product is used by on or nea...

Page 3: ...3 Remove all electro magnetic or RF generated equipment from close proximity to avoid electromagnetic interference 14 The Control Unit will have minor heat generated in operation avoid prolonged contact 15 When the main power supply is lost or has failed temporarily the Control Unit will stop working and the power failure alarm will sound for up to 20 minutes This is normal and the product will re...

Page 4: ...NUAL BEFORE USE KEEP AWAY FROM FLAMMABLE MATERIALS IP21 WATER AND DUST PROTECTION CLASSIFICATION FUSE SPECIFICATION DISPOSAL OF ELECTRICAL ELECTRONIC EQUIPMENT WEEE THIS PRODUCT SHOULD BE HANDED OVER TO AN APPLICABLE COLLECTION POINT FOR THE RECYCLING OF ELECTRICAL AND ELECTRONIC EQUIPMENT UL CERTIFICATION LOGO COMPLIACE WITH IEC60601 1 WITH RESPECT TO ELECTRICAL SHOCK FIRE AND MECHANICAL HAZARDS ...

Page 5: ...5 2 INTENDED USE 5 3 PRODUCT DESCRIPTION 5 4 PRODUCT INSTALLATION GUIDE 6 5 PANEL DISPLAY AND OPERATION GUIDE 8 6 CLEANING 13 7 STORAGE 14 8 MAINTENANCE 14 9 DISPOSAL OF AIR MATTRESS 15 10 TROUBLESHOOTING 15 11 TECHNICAL DATA 16 ...

Page 6: ...oxide 3 PRODUCT DESCRIPTION The Topaz 8 Plus Low Air Loss Mattress System is an alternating mattress replacement system used in the prevention and relief for patients with or vulnerable to pressure wounds The Topaz 8 Plus Low Air Loss Mattress System offers patients a comfortable and relaxing support surface by using the established principles of alternating therapy which can both prevent skin bre...

Page 7: ...raps or binding side straps 4 Ensure the CPR valve is at CLOSE position before turning on the power 5 Position the Control Unit by its elastic hanger brackets over footboard of the bed The elastic hanger brackets will self adjust onto the footboard tightly 6 Remove the Transport Cap of the hose connector and connect the hose connector to the Control Unit Firmly push the hose connector into positio...

Page 8: ...oltage CAUTION The Control Unit can only be used with the mattress recommended by the manufacturer Do not use the Control Unit for any other purpose WARNING Do not place the Control Unit in any area where power cord can come off easily 10 For patient transportation press Auto Firm button and wait for 5 minutes for the mattress to be inflated Disconnect the hose from the Control Unit and put on the...

Page 9: ...nction Mode Selection Alternate Static Constant Low Pressure Panel Lock out Comfort Control Auto Detection 5 1 1 Alarm Mute Press the alarm mute button to temporarily suspend alarms Should the situation not be resolved and fault conditions continue the alarm will resume notifying the patient and caregiver 5 1 2 Alternate Time Display Alternate cycle time can be selected from 10 30 minutes by press...

Page 10: ...ed cycle time until another mode is selected Static the mattress maintains at the selected pressure The Control Unit will automatically fall back to alternation mode after 20 minutes Constant Low Pressure CLP the mattress maintains at a reduced selected pressure in static mode 5 1 6 Panel Lock Out Press the Lock Out button to lock the panel Should the panel remain untouched for 30 seconds the Lock...

Page 11: ... then the Control Unit will enter operate mode directly P S To test the alarm battery press to turn off the power and the power failure alarm should be triggered Refer to 5 2 3 Audiovisual Alarm if the alarm is not triggered Press the Operate button and the system will begin to inflate and the Auto Firm indicator will be flashing The mattress should be fully inflated within 60 minutes and automati...

Page 12: ...e to turn off the system refer 5 2 4 Alarm Mute Table 1 Weight and Comfort Level Reference 5 2 2 CPR When CPR needs to be performed quickly rotate the CPR valve to OPEN position at the same time disconnect the hose connector from the Control Unit to speed up the air release 5 2 3 Audiovisual Alarm Power Failure When electrical shortages occur or power cord is unplugged without turning off the Cont...

Page 13: ...pressure alarm if the pressure resumes back to normal then the low pressure alarm will stop When more than one alarm is triggered the alarm will be performed according to priority level Refer to Table 2 Warning Code Reference Table for priority level Table 2 Warning Code Reference PRIORITY HIGH LOW WARNING CODE INDICATOR LED AUDIBLE OUTPUT MODE CONDITION OF OUTPUT WARNING DESCRIPTION REMARKS 0 N A...

Page 14: ...ay from dust If other detergent is used choose one that will have no chemical effects on the surface of the plastics case of the Control Unit CAUTION Do not immerse or soak the Control Unit Clean the mattress cover by using single use wipes with a solution of neutral detergent and hand hot water Rinse thoroughly with clean water and damp dry the mattress using single use wipes Disinfecting the cov...

Page 15: ...based product for cleaning CAUTION After cleaning dry the mattress without direct exposure of sunlight 7 STORAGE To quickly vacuum air out from mattress for storage rotate the CPR valve to OPEN position and disconnect the hose connector to release the air Lay the mattress out flat and upsides down Roll from the head end towards the foot end Packing strap can then be stretched around the rolled mat...

Page 16: ...tches with the Control Unit model xxBBB xxx The AAA should be the same as BBB If not please contact your dealer Check if the Transport Cap is removed and make sure the connector is not broken The Control Unit is not working Check if the plug is connected to the mains supply Check if the power switch is switched to ON position press Check if there is a blown fuse Power failure alarm failed If the C...

Page 17: ...Table 1 Weight and Comfort Level Reference Table and wait for a few minutes for better comfort Follow the procedures The low pressure light is constantly flashing and the alarm has sounded for inspection If the above information does not solve the problem please contact your local dealer or agent for further support 11 TECHNICAL DATA 11 1 Product Specification CONTROL UNITUNIT AIR MATTRESS DIMENSI...

Page 18: ...IONS OPERATION ENVIRONMENT 5 40 15 RH 93 RH no condensation STORAGE ENVIRONMENT 25 70 93 RH no condensation ENVIRONMENT PRESSURE 70 kPa 101 3kPa ENVIRONMENTHORIZONTAL LEVEL 3000m WATER AND DUST PROTECTION CLASSIFICAITON IP21 ...

Page 19: ...sions CISPR 11 Group 1 The device s uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The device s is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that su...

Page 20: ... s to earth 1kV differential mode Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 s 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip i...

Page 21: ...nsmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE1 At 80 MHz and 800 MHz the higher frequency range applies NOTE2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection f...

Page 22: ...on distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 𝑃 80 MHz to 800 MHz d 1 2 𝑃 800 MHz to 2 5 GHz d 2 3 𝑃 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the t...

Reviews: