User Manual – Diamond 8 Plus
Diamond 8 Plus
Alter
Micro Low Air Loss Pressure Relief System
User Manual
Manufactured by: Moxi Enterprises, LLC.
Distributed by: Quart Healthcare Inc.
www.quarthealthcare.com
Page 1: ...ser Manual Diamond 8 Plus Diamond 8 Plus Alternating Micro Low Air Loss Pressure Relief System User Manual Manufactured by Moxi Enterprises LLC Distributed by Quart Healthcare Inc www quarthealthcare com ...
Page 2: ...cold conditions Operation Temp 5 C 40 C R H 15 93 During operation cell phone with a maximum output power of 2W shall be kept at least of 3 3 m The plug is used as the disconnection device from the AC mains The plug shall be located for easily accessible No modification of this equipment is allowed Contraindications The system should not be applied to patients suffering from polytrauma with fractu...
Page 3: ...2 inches of enclosed convoluted foam to provide extra protection and comfort for the patient in case of power failure and the mattress deflates Control Unit Features pump 5 digital pressure setting for individual preference The pump is set to default to Alternation mode for daily operation Press to select Static mode if required In addition the caregiver can select the Static Function stopping the...
Page 4: ...h Canada Class 2 Mode of Operation Continuous IP classification IP21 First number 2 describes the level of protection from solid objects up to 12 mm e g person s fingers Second number 1 describes the level of protection from liquids Protection against vertically falling drops of water e g condensation Mattress Features Part No FM QHI0002 Device Identifier Health Canada FM MOX0003S Size cm 89 W x20...
Page 5: ...d iii Secure straps beneath the mattress to the bed frame iv Hang the control unit on the foot board of the bed frame v Attach the air tube connectors to the socket on the left panel of the control unit vi Verify that air hoses are not kinked under the mattress vii Attach cover to mattress viii Plug in the control unit and turn on the master power switch on the right side panel the STANDBY LED wil...
Page 6: ...ks or loose tubing connection result in low pressure Lock Function Press the LOCK button for 3 seconds to unlock the face plate Static Function Push the STATIC button and adjust the comfort control by pressing the SOFT FIRM button to achieve maximum patient comfort On this mode the system provides only low air loss therapy Perform a hand check by placing hand under the patient s buttocks between t...
Page 7: ...ace with clean covers Covers can be washed and thermally disinfected in a washing machine following below procedure Never use phenol based cleaning solutions Industrial Break washes Cold 10 minutes Main washes 60o C 6 minutes Main washes 70o C 10 minutes Extraction 2 minutes 3 Cold Rinses Extraction 5 minutes Domestic Pre wash Cold Main Wash 70o C 10 minutes Extraction 2 minutes Cold Rinses Extrac...
Page 8: ...se phenol based cleaning solution The air filter should also be cleaned and checked as often as possible at a minimum of every six months Air Filter can be removed by pinching center of the filter and pulling outward from the back of the Therapy control unit Cleaning procedure for air filter 1 Remove Air filter and Replace a new Filter 2 Use a soft bristle brush to remove dust and difficult dried ...
Page 9: ... foot of the bed Use the base mounted straps to secure Place the system in a plastic bag for storage It is recommended that the following guidelines are used whenever the system is being stored or transported to another location Temperature limitations 2560o C C 70o C Relative humidity 30 93 4 Waste Disposal This Product has been supplied from an environmentally aware manufacturer that complies wi...
Page 10: ... low air pressure while the air pressure for other air cells is normal Is the connection between air cells and the manifold kinked Is there any air leakage from the air cells Adjust the connection between cells and manifold Replace with a new air cell 6 Warranty Quart Healthcare Products guarantees that this equipment is free from defects in materials and workmanship Our obligation under this warr...