Model QT-710/720 Radiographic Table
Chapter 3 Assembly & Installation
19
Table Alignment (Model QT-720 Only)
The Radiographic Table’s image receptor must be properly aligned with the pro-
jected light field of the collimator in order to attain accurate x-ray exposures.
This procedure may require slight repositioning of the table base, which may
affect the position or arrangement of the leveling shims, requiring an additional
check prior to final table mounting. Collimator calibration, tubestand/collimator
alignments, and collimator light field to x-ray field adjustments, as described in
their respective manuals, must be completed prior to performing this proce-
dure. If the tubestand to be used provides transverse travel of the x-ray tube
relative to the table, set the tube transverse travel in the center detent position
prior to performing this alignment.
1. Position the table under the collimator, in the approximate location where it
will be mounted.
Figure 12. Image Receptor/Collimator Alignment
Summary of Contents for QT-710
Page 4: ...Model QT 710 720 Radiographic Table THIS PAGE INTENTIONALLY LEFT BLANK ii Revision History...
Page 6: ...Model QT 710 720 Radiographic Table Table of Contents THIS PAGE INTENTIONALLY LEFT BLANK iv...
Page 12: ...Model QT 710 720 Radiographic Table x Warranty Information THIS PAGE INTENTIONALLY LEFT BLANK...
Page 13: ...Chapter INTRODUCTION 1 1...
Page 14: ...2...
Page 17: ...5 Chapter SPECIFICATIONS 2...
Page 18: ...6...
Page 23: ...11 Chapter ASSEMBLY INSTALLATION 3...
Page 24: ...12...
Page 33: ...Chapter OPERATION 4 21...
Page 34: ...22...
Page 39: ...Chapter MAINTENANCE 5 27...
Page 40: ...28...