CARE AND MAINTENANCE
When to See Your Quantum Rehab Provider for Service
The following symptoms could indicate a serious problem
with your power chair. If necessary, contact your Quantum
Rehab Provider. When calling, have the model number,
serial number, nature of the problem, and the error code if
available.
Motor noise
Frayed harnesses
Cracked or broken connectors
Uneven wear on any of the tires
Jerky motion
Pulling to one side
Bent or broken wheel assemblies
Does not power up
Powers up, but does not move
Loose seat or seating components
Corrective Maintenance
If the battery condition meter does not light up when you turn
on the power:
Check the harness connections. Make sure they are
tight.
Check the circuit breaker. Reset it if necessary.
Check the battery connections.
If the above conditions prove normal, you can load test the
batteries with a battery load tester. These testers are available
at automotive parts stores. Disconnect both batteries before
load testing and follow the directions that come with the load
tester. If either one of the batteries fails the load test, replace
both of them. If your power chair still does not power up,
contact your Quantum Rehab Provider.