OPERATING INSTRUCTIONS

This chapter presents detailed instructions on the operation of the 9700.

Overview

The 9700 provides digitally controlled, single shot, current pulses.  The current is
provided by a 3700mF capacitor bank charged to approximately 44V.  The ability of
the instrument to maintain the requested current will be load dependent, limited by
the voltage and the stored energy.  Thus the output pulse is energy limited by the
amount of energy stored on the capacitor bank (CV

2

/2 = 36.2 Joules).  Long pulses or

large loads will result in droop at the end of the pulse as the capacitor bank
discharges.  The table below shows the maximum droop free pulsewidth as a function
of resistance and current.

--

--

300

860

10

--

90

425

990

5

50

160

500

1,060

2

75

190

525

1,090

1

10

5

2

1

RESISTANCE (ohms)

CURRENT (A)

MAXIMUM CONSTANT CURRENT PULSEWIDTH (ms)*

The current in the Model 9700 is controlled by a high-side power mosfet.  Care
should be taken when making connections.  A switch is provided for each channel to
disconnect the load from the capacitor bank.

The 9700 uses a single-line LCD to display menu selections.  The front panel keys
are used to “navigate” through sets of menus to set various options and control the
current pulse generator.  Some keys are only active at certain times or have different
behavior depending on what state the menus are in -- once understood, the menus can
be quickly and easily manipulated.  The 9700 also contains non-volatile memory for
storing up to twelve (12) configurations, so you need not go through the entire setup
process each time the instrument is used.

9

Summary of Contents for 9700 Series

Page 1: ...Quantum Composers Inc PO Box 4248 Bozeman MT 59772 Phone 406 582 0227 Fax 406 582 0237 www quantumcomposers com 9700 Series Pulse Generator User s Manual...

Page 2: ...Interlock Toggle Switch 7 Sync Out 7 External Trigger 6 Output Sections 6 Run Button 6 Stop Button 6 Right Button 6 Left Button 5 Down Button 5 Up Button 5 Menu Button 5 Enter Button 5 Control Button...

Page 3: ...and Protocol 16 Command Language 15 GPIB 15 Serial RS232 Pinout 15 4 Computer Interface 13 GPIB Addr optional 13 Volume 13 Fmark 13 Shot Counter 12 RECALL Configuration 12 STORE Configuration 12 Store...

Page 4: ...Table of Contents iii...

Page 5: ...loys a flexible microprocessor controlled architecture the 9700 can be customized to fit your exact requirements contact Quantum Composers for details Technical Support For questions or comments about...

Page 6: ...ge and those that will steer current away from your heart in case of accidental contact with a high voltage You will significantly reduce the risk factor if you know and observe the following safety p...

Page 7: ...transformer etc any time the equipment is connected to an AC outlet even if the equipment is turned off 6 Never work alone Someone should always be nearby to render aid if necessary Training in CPR f...

Page 8: ...4...

Page 9: ...menu selections and settings The left half of the display shows the current menu selection while the right half shows the current setting of the item associated with that menu Control Buttons Item But...

Page 10: ...ge and units for the current item Stop Button Disables output of pulses on all channels Run Button Enables output of pulses Output Sections There are two 2 output sections on the front of the 9700 eac...

Page 11: ...ay parameter Interlock A switch is provided to arm disarm the system This switch is used with the interlock input on the rear panel to control pulse generation of the system A READY indicator is provi...

Page 12: ...8 interface is available as an option with the 9700 AC Input A standard EIC instrument type cord should be used The 9700 is supplied with a power cord that fits the type of socket used in your country...

Page 13: ...and current 300 860 10 90 425 990 5 50 160 500 1 060 2 75 190 525 1 090 1 10 5 2 1 RESISTANCE ohms CURRENT A MAXIMUM CONSTANT CURRENT PULSEWIDTH ms The current in the Model 9700 is controlled by a hi...

Page 14: ...e pressing and holding the Left arrow key will display a brief description of the current item When the right most digit is active pressing and holding the Right arrow key will display the range and u...

Page 15: ...s from 00 00 to 25 00A may be selected with a resolution of 10mA Current is provided from a 3 700mF capacitor bank which is charged to nominally 44V The maximum current available will therefore be lim...

Page 16: ...cation one 1 twelve 12 Note If you overwrite a particular location its previous contents are destroyed A configuration is stored by using the Up and Down arrow keys to select the memory location numbe...

Page 17: ...to zero the counter If you press any other key the shot counter will not be reset Fmark The character used as a decimal point can be set to either a period or a comma The period is generally used in...

Page 18: ...14...

Page 19: ...nnected to pin 6 5 Ground 6 DSR connected to pin 4 7 RTS connected to pin 8 8 CTS connected to pin 7 9 No connection The voltage level of the Transmit signal is 10V the Receive signal should be at lea...

Page 20: ...ntified the pulse generator returns an ok followed by a carriage return and a line feed ASCII 13 ASCII 10 otherwise it returns a 1 a question mark followed by the number one a carriage return and a li...

Page 21: ...g Echo Enabled MODE cr lf 100 cr lf Echo Disabled 100 cr lf Communication Protocol Communication parameters are set at 9600 baud 8 bits parity none and 1 stop bit To help establish RS232 communication...

Page 22: ...s the delay in microseconds 10 for Channel 2 0 999999998 T2 Dly Sets the pulse width in microseconds 10 for Channel 2 100 100000 T2 Wid Sets the current level in 10 s of mA 100 1A for Channel 2 0 2500...

Page 23: ...TION 200ns P PULSEWIDTH 100 s to 100ms P RESOLUTION 200ns P ACCURACY 100ns 0001 x delay P TIMEBASE 5MHz 25PPM crystal oscillator P RMS JITTER 100ns P TRIG DELAY Ext Trig to To 10 s EXTERNAL TRIG GATE...

Page 24: ...5A per channel P DC CURRENT LIMIT 150mA each channel COMPUTER INTERFACE P RS232 9600 Baud All instrument functions and settings may be controlled over the interface bus GENERAL P STORAGE Twelve 12 com...

Page 25: ...entry module To replace fuses follow the instructions given below 1 Unplug the 9700 from power source and remove power cord from the back panel 2 Remove the fuse holder by pushing down on the center t...

Page 26: ...22...

Page 27: ...6 CUSTOM FEATURES Long Pulse Option Specifications 10 s 100ms 10A for the duration of the width 23...

Reviews: