Quantum Dynamo
www.quantumrehab.com
43
Drive Program Select Key
The drive program select key enables you to select a drive program and an actuator mode. Your Dynamic DX controller
was preprogrammed at the factory for five drive programs— 1 (slowest) to 5 (fastest). The drive mode is indicated by a
number in the center of the keypad.
NOTE: The drive mode settings are preset at the factory. If your Quantum Rehab Specialist changes these
settings, please make note of these changes.
To change the drive mode program:
1. Press the drive mode select key.
2. Move the joystick right to increase the drive mode program.
3. Move the joystick left to decrease the drive mode program.
4. Press the drive mode select key.
Drive Mode and Actuator Display (also Remote Status Display)
This is a seven-segment light that displays the drive program. It also displays a drive inhibit and actuator mode (if appli-
cable), when they occur.
Actuator Select Keys with Lights
If your power chair is equipped with power accessories such as a power seat or power elevating leg rests, you can operate
them through the remote keypad. Contact your Quantum Rehab Specialist for information on how to operate these acces-
sories.
Remote Status Light
This light will flash if there is an internal Dynamic DX Remote fault, or if an OONAPU fault has occurred. See “Out Of
Neutral At Power Up.”
Horn Key
The horn key activates the horn.
Battery Saver Feature
When the battery capacity drops to below 21V (typically two lights), the controller will reduce power chair performance
to conserve battery power.
Off-board Charger/Programming Socket
You may use an off-board charger to charge the power chair batteries through the 3-pin socket located on the front of the
Europa. See figure 29. If you use an off-board charger, the charger current should not exceed 8 amps. Contact your
Quantum Rehab Specialist for more information.
NOTE: The socket may also be used for reprogramming the Europa. Contact your Quantum Rehab Specialist
for more information.
NOTE: Changes to the programming can only be made by the manufacturer, an authorized representative of
the manufacturer, or a Quantum Rehab Specialist.
V I I . O P E R A T I O N