TPA-MHL-8R - Quick Start Guide
68-00224 - Rev. A4
Page 17
August 31, 2011
Use the procedure below to make the connections the TPA-MHL-8R for running source
compliance tests on an MHL source device.
1. Apply power to the TPA-MHL-8R using the power adapter provided.
2. Make the connections as shown in the diagram below. For this application you will be using
the section on the TPA-MHL-8R labeled
980 CONNECTIONS
– SOURCE DUT.
a. Connect an MHL cable from the MHL source device under test to the MHL IN FROM
DUT connector on the TPA-MHL-8R.
b. Connect an HDMI cable from the OUT TO 980 PORT 1 connector of the TPA-MHL-8R
to the Rx port on the 980 Protocol Analyzer.
3. Refer to the 980 MHL Compliance Test User Guide for details on how to run the MHL
compliance test on an MHL source.
Connect MHL
cable to MHL
mobile device
(MHL source)
Connect HDMI
cable to Rx Port
on 980 Protocol
Analyzer