SERVICE MANUAL
CAUTION:
Disconnect the power cord before any intervention on the instrument.
5-2
FUSE REPLACEMENT
Fuses specifications:
•
Glass fuses
•
1 AT - 250V
•
5 x 20 mm
With a screwdriver pull the
main fuses block at the
rear of the Aviso.
5- MAINTENANCE
5-3
A-PROBE CALIBRATION
Biometry and ProBeam probes can be installed and used on the 3.0 Aviso
software, without doing a calibration between each use of the probe.
A calibration can be performed for each probe, and permits to get the
correct position of the first marker in Contact.
Important:
After installing the new software version, the calibration of the
probes is not saved (due to the modification of the biometry and
probeam controls), so you have to perform the biometry calibration
for each probe: Biometry and ProBeam probes. Before to upgrade,
you can note the zero parameters and the cursor position, then
perform the upgrade and insert the noted parameters, but you have
to check the calibration is correct.
Procedure:
• Run
« Aviso_Setup.exe »
software, in the
« C:\Program Files\Quantel
Medical\Aviso »
directory.
• Click on
« Setup »
and select
« Probes »
tab,