CIB418-01
66
IEC
61000-
4-2
kV,
±
15
kV
kV,
±
15
kV
should be at least
30%
Electric
al fast
transie
nt/burst
IEC
61000-
4-4
+ 2kV for
power
supply
lines
+ 1kV for
input/outp
ut lines
+ 2kV for
power
supply lines
Not
applicable
Mains power
quality should be
that of a typical
home healthcare
environment.
Surge
IEC
61000-
4-5
+ 0.5kV,
+1kV
line(s) to
line(s)
+ 0.5kV,
+1kV,+
2kV line(s)
to earth
+ 0.5kV,
+1kV
line(s) to
line(s) Not
applicable
Mains power
quality should be
that of a typical
home healthcare
environment.