3
For Research Use Only
the tray deck and cause dust particles to be visible in
your image. Please take precautions to keep the lab
bench surface and floors free from dust.
Space
The footprint of the Q-View
Imager LS is 25.4 W x 23.5
D x 55.88 H cm (10 x 9.25 x
22 inches). Ensure the space
around the Q-View Imager
LS is free of chemicals that
can spill or other equipment
that can fall and damage the
Q-View Imager LS.
Vibration
The exposure times of the Q-View Imager LS can be
up to 5 minutes. It’s important that the location of the
Q-View Imager LS is free of excessive vibration. Make
sure equipment that might cause vibrations such as
vortex mixers, shakers, centrifuges, etc. are not co-
located on the bench with the Q-View Imager LS. Also
consider where the Q-View Imager LS is placed in
the laboratory. For example, the Q-View Imager LS
should not be placed on a bench near a door that
could be slammed or on a bench near a desk where
someone could bump it during exposures.
Q-VIEW IMAGER LS SETUP
Assembly
The Q-View Imager LS weighs approximately 9.12 kg
(20.12 lbs). To correctly setup the Q-View Imager LS
the following steps should be followed:
1.
Place the Q-View Imager LS in a clean work area.
Ensure the location meets the requirements
addressed in Placement in the Laboratory ()
2. Connect the wall cord to the Q-View Imager LS..
3. Plug the wall cord into a surge protector.
4.
Connect the USB cord to the computer that will