Model WBP-101 Wireless Pulsewave Blood Pressure Monitor
Model WBP-101 Wireless Pulsewave Blood Pressure Monitor
RECHARGE THE BATTERY
1. When the screen displays
"
LOP
",
it indicates that internal battery is
depleted.
P
lease recharge the unit immediately using the power adapter;
2. Press the button "
" to turn on the device, the screen should show
steady
“-0-“. If the "-0-" flashes, the battery is low of charge. You may
proceed with the current measurement, and recharge the unit in time.
3. During charging process, the charging indicator light is in orange; when
the battery is fully charged, the charging indicator turns into green.
CAUTION
• Please use the included power adapter or compatible power
supply equipment (
DC 5V
) to recharge the internal battery.
• During the process of charging, please do not use the product
;
please do not plug in the power adapter with wet hands.
• If the product is not to be used in 3 months, please store the
device with the battery fully charged.
SETTING UNITS
1.
While the device is powered off
,
press and hold the button
"
'
"
for over
five seconds to enter
“unit setting” mode. The screen should display “-U-“,
as shown below. If it reads
"
mmHg
"
at the top
, the
current unit is in
"
mmHg
". Similarly,
"
kPa" indicates current unit in
"
kPa".
2.
The unit display only lasts about two seconds. Then the device
goes into “idle” mode (displaying “-0-“).
3.
If you need to reset the unit, you must turn off the device by
pressing the button
"
'
" twice, t
hen repeat the above steps.
NOTES:
When in idle state
(displaying “-0-“ on the screen)
,
press and hold the
button
"
"
for about 5 seconds, the device will enter into
"
P mode
",
which is not intended for normal users
. To exit and resume to normal
operation, you can do one of the followings: a. Press and hold the button
"
"
for about 5 seconds
;
b. Wait for 15 to 20 seconds to automatically
exit
.
Calibration should be performed every three years. It is
recommended that the product be sent back to Qualsig Inc. or
authorized distributors for calibration. Otherwise, accuracy may goes out
of specification.
7
8