Table 3-6 lists the Error Messages for the IEEE-488 interface of the 1855 instrument. In
response to the command “SYSTem:ERRor?”, the 1855 unit responds with the error message
number and an error message string of up to 80 characters in length.
Table 3-6: Error Messages
Code Type
Message
0
No error
There is no error in the error queue
-102
Syntax error
Invalid character exists in the command string.
-104
Data error
Parameter is not defined in the command string.
-106
Illegal parameter
Parameter is not a valid command.
-202 Conflicting
Settings
Command
conflicts with instrument settings.
Example: Send ‘Trigger’ when mode is external.
-203
Data range
Data exceeds the valid range.
-211
Data stale
No resent measurement result.
Example: Send ‘Read?’ when in Standby status.
-224
Self-Test failed
Self-test via remote interface (*TST) failed.
-225
Excess errors
The error queue is full (more than 20 errors).
Queue cleared after power down or *CLS command.
-226
Query interrupted
Device status changed after query sent.
Output buffer will be cleared