Page 42 of 76
Interface
Figure 2.11 illustrates the two pages of parameters that can be programmed within the MEAS
DISPLAY for a SEQUENCE TEST. Figure 2.12 illustrates the two pages of parameters that can
be programmed within the MEAS DISPLAY for a STEP TEST. The two tests have the similar
programmable parameters with the exception of Charge Time, Range Dwell and Trigger. All
programmable parameters are explained in Paragraphs 2.5.1 through 2.5.8.
2.5.1
Test Voltage
The test voltage can be programmed from 1.00V to 650V. In MEAS DISPLAY press [F1] =
TEST V so that the 1.00 V box is highlighted. Use the up arrow or down arrow keys to
in/decrease the voltage in multi-V increments. The left and right arrows will increase/decrease
the voltage in 1V increments. The instrument default setting is 1.00V.
UP arrow [
⇑
] key: 6.3
→
10.0
→
16.0
→
25.0
→
35.0
→
50.0
→
63.0
→
100.0
→
160.0
→
200.0
→
250.0
→
350.0
→
400.0
→
450.0
→
500.0
→
550.0
→
600.0
→
630.0.
DOWN arrow [
⇓
] key: 630.0
→
600.0
→
550.0
→
500.0
→
450.0
→
400.0
→
350.0
→
250.0
→
200.0
→
160.0
→
100.0
→
63.0
→
50.0
→
35.0
→
25.0
→
16.0
→
10.0
→
6.3.
RIGHT arrow [
⇒
] key: increase voltage in 1V increments.
LEFT arrow [
⇐
] key: decrease voltage in 1V increments.
1.0
0.5mA
NEXT PAGE 1/2
2uA
Lc :
< MEAS DISPLAY: SEQ. TEST >
A
TEST V.
C.C.
RANGE
:
:
:
DISCHARGE
CHARGE
TEST
Vm = 0.0V
V
Select Test V
F1
1.0, 6.3 - 630V
630V - 6.3V
630V - 650V
6.3 - 1.0V
2.5.2
Constant Charge Current
The test current can be programmed from 0.5mA to 500mA. In MEAS DISPLAY press [F2] =
C.C so that the 0.5 mA box is highlighted. Use the up arrow [
⇑
] key to increase the current or
use the down arrow [
⇓
] key to decrease the current in 5/50mA increments. The left and right
arrows will increase/decrease the current in 1mA increments. The instrument default setting is
0.5mA.
NEXT PAGE 1/2
2uA
Lc :
< MEAS DISPLAY: SEQ. TEST >
A
TEST V.
C.C.
RANGE
:
:
:
DISCHARGE
CHARGE
TEST
Vm = 0.0V
1.0 V
0.5 mA
F2
0.5 - 450.5mA
450.5 - 0.5mA
450.5 - 500.0mA
500.0mA - 450.5
NOTE:
For the WV test, the range of C.C. is from 0.5mA to 150mA.