7 . 2 C L E A N I N G , D I S I N F E C T I O N A N D C A R E
For cleaning and disinfection of the device and its accessories there should not be used any agents containing
higher portions of phenol derivatives, alcohol, compounds of chlorine or peracetic acid. It is recommended
to use disinfectants on aldehyde basis.
We recommend to clean the device with a soft, damp cloth. For tenacious stains you may also apply mild
detergent used for delicate plastic surfaces.
The device is not suited for heat sterilization or for sterilization with gases.
!
ATTENTION
Before cleaning or disinfection unplug the power supply out of the socket!
The device is suited for wiping disinfection. It has to be observed that no liquids enter the device. In no case
shall the plug or socket get wet. For cleaning or disinfection the device may not be drizzled.
Control Unit, Applicators
Do not immerse in water! Use normal household cleaners for synthetic materials. Only wipe with a damp
cloth. Be sure to keep sharp, pointed objects away from mat applicator and pillow applicator.
When using the coil mat and coil pillow, please make sure that no sharp objects are pressed in. Do not
expose the system to direct sunlight and protect it from frost.
Placing Mat under a Mattress
The mat can be placed under a mattress. A woolen blanket should be placed between the mattress and
the mat in order to avoid undesirable bacteria or fungi caused by sweat.
33