6 . 1 . 1 E R R O R C O D E S D U R I N G T I M E R T H E R A P Y
If a timer therapy (activated by the user through a chip card or via timer programming) could not be performed
a fault report will appear on the display the next time the device is started. This report must be acknowledged
by pressing the Start/Stop-key.
For each possible time-controlled therapy – timer 1 and timer 2, therapy 1 through 4 (chip card) – the display
will show the following messages:
In the last stated message the
>>N<<
states the number of the timer or therapy,
>>xyz<<
represents an error
code, meaning the following:
E101
No applicator has been connected, or the applicator does not correspond to the applicator selected
in a chip card therapy.
´
Please connect an applicator respectively the one prescribed on the chip card.
E102
The applicator does not conduct current. The applicator or its connection cable may be defect, usually
caused by rough or inappropriate handling.
´
Please check if the applicator has been connected and tightened correctly. If available test another
applicator on the device’s corresponding connection. In case of visible (mechanical) defects please
contact your service partner or distributor.
E103
The therapy could not be started since the device was already in operation, ulteriorly at the
programmed starting time.
´
Please check the starting times, correct them, if required and repeat the therapy, if necessary.
E104
The therapy could not be started due to an error within the device (communication problem).
´
Please contact your service partner.
FAULT REPORT :
(PLS SEE MANUAL)
GO ON WITH START
TIMER N: Exyz
30