Page
7
of
18
MANVALTOL-KT2 Rev. A, 6-15-2010
4.2.3
Once set, secure the paddle or basket shaft in place as
recommended by the appropriate dissolution tester
manufacturer. Then raise the dissolution drive head and
carefully remove the gage.
4.3
Checking Paddle / Basket height
4.3.1
After setting the paddle or basket height. The height can be
checked by using the Go / No-Go Gage as outlined below.
4.3.2
With the dissolution drive head lowered to its normal
operating position, the smaller knob labeled “GO” must fit
in between the bottom of the paddle or rotating basket and
the bottom of the vessel. While the larger knob labeled
“NO-GO” must not.
4.3.3
If the “GO” knob does not fit or the “NO-GO” knob does,
the height needs to be adjusted. (See sections 4.2.1 -4.2.3)
4.4
Measuring Paddle Wobble
4.4.1
For Hanson and Distek tester
4.4.1.1
The paddle wobble can be measured by assembling the
dial indicator assembly, small indicator shaft and
wobble bracket assembly as shown in Figure #3.
Figure #3
Wobble Bracket
Assy.
Adjustable
Foot
Dial Indicator
Assy.
Small Indicator
Shaft