MANVALTOL-KT2 Rev. A, 6-15-2010
Figure #6
4.4.2.3
Repeat sections 4.4.1.3
& 4.4.1.4
4.5
Measuring
Basket Shaft Wobble
4.5.1
For Hanson and Distek
Testers
4.5.1.1
The basket shaft wobble
can be measured by assembling the dial indicator
assembly, small indicator shaft and wobble bracket
assembly as shown in Figure #3.
4.5.1.2
Place the wobble gage onto the top of the dissolution
vessel and move the adjustable foot in or out until
the assembly is secured to the inside of the vessel as
shown in Figure #3.
4.5.1.3
With the basket shaft secured, lower and position the
dissolution drive head so that the dial indicator probe
is located about 2cm above the top of the shaft hub
and adjust the location of the indicator assembly to
establish a pre-load of at least one revolution of the
Centering Ring
Guide Post