QHY4040
6
6. The following adjust tools is for temperature regular control. When set to Auto , you can set the target
temerpature. The QHY4040 has maxium -35C delta T below ambinet. You can set a suitable target
temperature according the enviroment temperature. Please reserver some degree to avoid the camera
working in just -35 delta T, otherwise when enviroment get higher , you can not get the target temperature
and cause the dark frame calibration has problems. When set Auto off, you can set the cooler power to
adjust the TEC power. The range is 0% to 100%. In this mode, the sensor temeperature is not regulated.
Just the power is constant
Before using the cooler, it is best to read the "how to protect the cooler" in the FAQ.