Entering calibration data
1.
Enter Calibration Standard into the text field and hit done on the keypad.
2.
Pour Calibration Standard into storage/calibration cup until sensors are fully submerged.
3.
Wait for the value to appear on lower right corner of dialog box. This typically takes about one minute
but could take slightly longer.
4.
Hit OK
5.
After this point the application will bring up another dialog box to calibrate to a second standard if
applicable (e.g., pH or turbidity). Follow the same steps above for the second standard. After successful
calibration of second standard the application will prompt for an optional third standard calibration for
pH; this is not necessary but is recommended for a more accurate data representation.