Rev: 01
P/N: 95343
Page | 10
3.3.2 Find another MP25
If this button is pressed after an MP25 or MP25T is connected it will disconnect the current device. This button
can also be pressed if a user is trying to connect to an MP25 or MP25T and the device doesn’t show up on the
list of Bluetooth devices.
3.3.3 View PurgeScan Logs
This button will bring up a dialog box allowing the user to view or download any PurgeScan Logs that have been
taken with the device. The
PurgeScan
Logs button located in the PurgeScan Menu is also another way to view
the save log files. In the upper left corner is the name of the log file, when pressed a list of other logs will appear
and can be viewed. These files will be downloaded onto the selected File Save Location, see Section 3.7.6 on
how to set the save location.
3.3.4 Settings
In the settings menu the user can control the PurgeScan Logging and Logs, Calibration of the sensors, Backup
Logs and app driven device settings.
3.4 Calibration and Calibration Logs
3.4.1 Calibration
Calibration of the sensors on the sonde is recommended each time the user goes out to sample for the day. The
calibration function on the sonde is located in the menu under Settings > MicroPurge Sonde > Calibrate. The
table below lists the parameters that can be calibrated on the MP25 and MP25T and the commonly used
methods and standards.