ACCURACY TEST
The instrument can be tested with the test kit comprising of:
-
The 10ppm cylinder
-
Flow regulator 1l/min.
Service and Calibration
The SEM5000 should be regularly serviced. QED recommends that the instrument is
calibrated on an annual basis.
NOTE: The next calibration date can be seen during the warm up sequence after
powering the instrument and also on a label on the underside adjacent to the battery
charging connector.
The hydrophobic internal filter should be checked periodically, (typically this will be
weekly during periods of heavy use).
Alarms and Error Information
The table below gives the different alarm conditions or information about errors.
Displayed Alarm or Error
Error
Conditions
Acoustic alarm and the
LED’s flash
GAS alarm
Concentration measured above the alarm
threshold.
The display shows the
message ERROR PUMP,
LED
’s flash
and the pump
icon is displayed with a
cross
Pump stopped
The pump is stopped possibly due to the ingress of
moisture or a high quantity of dust. -