Declaration – electromagnetic emissions and immunity for equipment and systems that are not life-supporting and are specified for use only in a shielded location
The QardioArm Wireless Blood Pressure Monitor declaration-electromagnetic immunity
The QardioArm Wireless Blood Pressure Monitor system is intended for use in the electromagnetic environment specified below. The customer or the user of the Wireless Blood
Pressure Monitor system should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment - guidance
Conducted RFIEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
N/A
Portable and mobile RF communications equipment should be
used no closer to any part of the EQUIPMENT or SYSTEM
including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the
transmitter. Interference may occur in the vicinity of equipment
marked with the following symbol
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3V/m