6.4 Venting (May 2020)
●
Xe-1 models are designed to release warm exhaust air directly into the room (See
Figure 6.4b,c,d
).
●
Xe-1 testers do not require a vent hood or ducting.
●
Do not attach to a powered vent.
Figure 6.4a: Heat load dissipation by model.
IMPORTANT:
Air intakes and exhausts must not be obstructed. Keep filters clean.
See
Section 16.5
for more information on filter cleaning.
Figure 6.4b: Q-SUN Xe-1-SE air intake and exhaust areas.
Q-SUN Xe-1-BCE, Xe-1-SCE
Figure 6.4c: Q-SUN Xe-1-BCE and Xe-1-SCE chiller cooling air intake areas.
Air Exhaust
Air Intake
Air Exhaust
Rear View
29
Section 6. Set Up
Model
Heat Load Dissipation
Q-SUN Xe-1-B, Xe-1-SE
1,200 watts (4,100 BTU/Hr)
Q-SUN Xe-1-BCE, Xe-1-SCE
2,400 watts (8,200 BTU/Hr)
Q-SUN Xe-1-WE
1,400 watts (4,800 BTU/Hr)
Q-SUN Xe-1-B, Xe-1-SE