background image

For further detail on laboratory environment requirements, please contact 

Q-Lab Repair and Tester Support. 

 

See 

Section 9 

for contact information.

11

Figure 4.2d: Keep tester away from 

sources of hot or cold air. 

Other Unsuitable Environments

 

Outdoors:

 Rain and dust will corrode or short out electrical components.

 

Metal Dust / Metal Chips:

 Do not locate the tester near metal cutting machines or metal grinding machines. 

Conductive metal dust or metal chips in the air will damage electronic components.

 

Carbon Fibers:

 Do not operate the tester where carbon fibers or carbon reinforced plastic are being cut. The 

conductive carbon fibers will damage electronic components.

 

Conductive Pigments:

 Do not operate the tester where carbon black or other conductive pigment dust is in the air. 

The conductive dust will damage electronic components.

 

Other Corrosive Gases:

 Do not expose the tester to acid fog, SO

2

 gas, or other corrosive gases.

 

Excessive Voltage:

 The electrical supply to the tester must be no more than 10% higher than the voltage listed on the 

nameplate.

 

Low Voltage:

 Recurring “brown-outs” or voltages less than 90% of the rated voltage will damage electrical 

components.

 

Water Leaks from Ceiling:

 Water leaking onto the tester will damage electrical components.

X

HVAC Vent Directly 

Above Tester

Section 4. Operating Environment

Summary of Contents for Xe-1

Page 1: ...QUICK SET UP GUIDE Xenon Test Chamber Q SUN Xe 1 LX 5090 QS Revision Date 11 Apr 2022...

Page 2: ...ronments May 2020 10 5 Uncrating 12 5 1 Overview Feb 2022 12 5 2 Models Without Chiller Feb 2022 12 5 3 Models With Chiller Feb 2022 16 6 Set Up 22 6 1 Dimensions and Space Requirements Mar 2022 22 6...

Page 3: ...and or humidity faults Transportation and Storage Temperature 40 C to 80 C Installation Category Category II for transient over voltages Pollution Control Pollution Degree 2 Sound Pressure Level Sound...

Page 4: ...Symbols Aug 2020 Electrical Shock Hazard Hot Surfaces Hazard Attention Finger Hand Crushing Hazard Local waste recycling regulations per the WEEE Directive 2002 96 EC on Waste Electrical and Electron...

Page 5: ...of this product It is the responsibility of the user of this manual to establish appropriate safety and health practices and to determine the applicability of regulatory limitations prior to use If t...

Page 6: ...label Figure 2d Shock hazard warning label Heat and Electrical Shock Hazards Warning Labels All Q SUN Xe 1 models are equipped with a lamp access door interlock switch that turns off the xenon lamps...

Page 7: ...ght There is no Ultraviolet UV hazard from any Q SUN tester in normal operation with the doors closed The Q SUN lamps produce UV light that can cause severe sunburn eye inflammation and damage to your...

Page 8: ...rmation on the installation operation and maintenance of Q SUN Xe 1 B testers Q SUN Xe 1 Models Covered in This Guide Model Configuration Installation Control Interface Xe 1 B B Basic Tester Tabletop...

Page 9: ...relative humidity RH for Q SUN testers is 23 5 C and 50 25 RH Operating outside the recommended range or in rare cases even within it certain standards or test cycle conditions may not be achievable...

Page 10: ...a room with machines or processes that generate dust particles vapors gases etc Figure 4 2b X Figure 4 2a Do not install testers in a room with corrosion chambers Figure 4 2b Do not locate testers in...

Page 11: ...carbon reinforced plastic are being cut The conductive carbon fibers will damage electronic components Conductive Pigments Do not operate the tester where carbon black or other conductive pigment dus...

Page 12: ...g 5 Uncrating 5 1 Overview Feb 2022 Carefully read these instructions before uncrating the tester Follow all local OSHA EHS and other applicable equipment operation and material handling safety requir...

Page 13: ...Packaged with wooden frame continue with Step 3 Packaged with carton only go to Step 7 3 Locate the Phillips screws in the bottom front and back horizontal boards 4 Remove the screws Screws in Back Sc...

Page 14: ...r Use Fork Lift to Remove Crate in One Piece 7 If there is an accessory package cut the band and remove the plastic wrap Set the package aside Accessory Package Cut 8 Cut the bands holding the carton...

Page 15: ...that at least two people remove the Q SUN from the skid 12 Carefully lift the tester off of the skid To complete tester installation see Section 6 1 through Section 6 4 13 Go to the sections listed ab...

Page 16: ...crating and setting up the tester are located in the envelope shown in Figure 5 3d Figure 5 3a This packaging has a carton banded to a wooden skid Carton Skid Open First Label Figure 5 3b This packagi...

Page 17: ...ps screws in the bottom front and back horizontal boards 4 Remove the screws Screws in Back Screws in Front Remove 3 Remove 3 5 If a fork lift is available use it to carefully lift the wooden frame up...

Page 18: ...ding the carton to the skid Remove the bands and boards Cut 9 Carefully lift the carton up and off of the tester 10 Remove all packaging materials from inside and outside the tester Remove Air Filter...

Page 19: ...r filter frame points up Close the door 16 Screws 2 each side attach runners to skid Screws Runner Runner One Screw in Right Rear 17 Remove the four 4 screws Remove 4 At least two people are needed to...

Page 20: ...unner use a short screwdriver to remove the screw holding the runner to the unit Remove Screws 22 Slide the unit the opposite way until it overhangs the skid about 100 mm 4 Slide Backward 100 mm Off S...

Page 21: ...ling glides do not hit floor drain cracks etc as the glides may be damaged If specified at the time of order the chiller may be equipped with casters Leveling Glides To complete tester installation se...

Page 22: ...65 cm 39 0 28 25 26 External Dimensions With Casters W H D No Casters 78 cm 143 cm 79 cm 30 5 56 1 31 0 No Casters The Q SUN Xe 1 B and Xe 1 SE are designed for tabletop or benchtop installation Figu...

Page 23: ...79 cm 91 cm Vent Figure 6 1b Q SUN Xe 1 BCE Xe 1 SCE minimum clearance and space requirement Q SUN Xe 1 WE is designed for tabletop or benchtop installation Figure 6 1c shows external dimensions and s...

Page 24: ...IS NOT SUPPLIED with Q SUN Xe 1 testers The user must supply and attach a properly sized plug to connect the Xe 1 to the Main Power Supply Q SUN Xe 1 testers are provided with a 3 conductor 14 AWG 4 9...

Page 25: ...r system shown in Figure 6 3a produces water pure enough for spray systems This type of RO DI system is required for Xe 1 SE Xe 1 SCE and Xe 1 WE testers Important Use Type I not Type II anion in the...

Page 26: ...solids colloidal silica organic and biological contaminates 0 2 0 5 M cm Rough purification stage D Ultraviolet Lamp Disinfect water Rough purification stage E Mixed Bed Tank Final polishing to remov...

Page 27: ...ied with the tester Refer to the diagrams included with the tester or visit Q Portal for detailed plumbing information 27 Figure 6 3c Hose barbs and elbows supplied with models requiring supply and dr...

Page 28: ...e To Chiller Drain Hose Figure 6 3j Chiller drain hose reducer Rear View 1 3 cm 1 2 Drain Hose For Raised Drain Installations If a drain hose must rise more than 100 mm 4 0 above the tester drain outl...

Page 29: ...e obstructed Keep filters clean See Section 16 5 for more information on filter cleaning Figure 6 4b Q SUN Xe 1 SE air intake and exhaust areas Q SUN Xe 1 BCE Xe 1 SCE Figure 6 4c Q SUN Xe 1 BCE and X...

Page 30: ...reas Q SUN Xe 1 WE Figure 6 4e Q SUN Xe 1 WE air intake and exhaust areas Air Intake Air Exhaust Air Exhaust Rear View Lamp Ballast Cooling Air Intake Lamp Ballast Warm Air Exhaust Chiller Excess Cool...

Page 31: ...men tray pulled out and specimens placed on tray 7 2 Q SUN Xe 1 All E Models Control Panel May 2020 The power switch is located on the tester control panel Figure 7 2a Two touch screen displays provid...

Page 32: ...7 2b Control interface and test status display Figure 7 2c Menu screen Figure 7 2d Status screen Black Panel 70 68 Main Menu Settings Diagnostics Contact Q Lab Manage Cycles Manage Test Timers Calibra...

Page 33: ...ges Figure 7 3c The test monitor display shows test setup parameters and provides real time data on actual test conditions Figure 7 3d Figure 7 3a Power switch on Xe 1 B control panel Program Escape E...

Page 34: ...mited to the purchase price paid Damage due to accident or abuse is not covered Labor and travel costs are not covered Q Lab Corporation makes no other warranties including implied warranties of merch...

Page 35: ...office time Please contact the nearest Q Lab international office by phone or email see contact information below You can also visit our website at www q lab com to register your tester to access add...

Reviews: