![pyroscience Pico-pH-SUB Manual Download Page 23](http://html1.mh-extra.com/html/pyroscience/pico-ph-sub/pico-ph-sub_manual_1632886023.webp)
Pico-pH-SUB
| Manual
© PyroScience GmbH
23
N
Number of optical channels. For the Pico-pH this value is 1.
R
Firmware version, e.g.
R
=403 designates firmware version 4.03
S
Bit field about available sensor types and supported optical analytes as
follows:
Bit 0-7: Available Sensor Types
Bit 8-15: Supported Optical
Analytes
Bit 0: optical channel(s)
Bit 8: oxygen
Bit 1: sample temperature (typ.
Pt100)
Bit 9: optical temperature
Bit 2: pressure
Bit 10: pH
Bit 3: humidity
Bit 11: CO2
Bit 4: analog in
Bit 12: reserved
Bit 5: case temperature
Bit 13: reserved
Bit 6: reserved
Bit 14: reserved
Bit 7: reserved
Bit 15: reserved
Example:
S
= 1 + 2 + 4 + 8 + 32 + 1024 = 1071 means, that the device
provides an optical channel as well as sample and case temperature,
pressure, and humidity sensors, and the optical channel supports the
analytes pH.
B
Firmware build number starting at 1 for each firmware version (reflects
minor firmware revisions which normally do not require a software or
firmware update for the user)
F
Bit field about available features as follows:
Bit 0: analog out 1
Bit 5: battery
Bit 1: analog out 2
Bit 6: stand-alone logging
Bit 2: analog out 3
Bit 7: sequence commands
Bit 3: analog out 4
Bit 8: user memory
Bit 4: user interface (display,
buttons)
Bit 9-31: reserved
Example: F = 1 + 2 + 4 + 8 + 256 = 271 means that 4 analog outputs are
supported and the module possesses a user memory. Note, the optional
analog outputs require additional hardware (more information on request).
Example Communication:
Command:
#VERS
˽
1
↵
Response:
#VERS
˽1˽
4
˽
403
˽
1071
˽
2
˽
271
↵
5.4.6
#IDNR
–
Get Unique ID Number
This command returns the unique identification number of the respective device.
Command:
#IDNR
↵
Response:
#IDNR
˽
N
↵
Output Parameters:
N
Unique ID number. Note, this parameter is given as an unsigned 64 bit
integer!
Returns the unique identification number of the device (does NOT correspond to the
serial number of the device).