background image

2

 

Introduction to the FireSting pro Meter 

The 

FireSting  pro 

comes  with  1,  2,  or  4  channels  (optical  sensor 

ST-connectors  1  to  4)  for  up  to  4  fiber-optic  sensors  and  one 
connector (T) for an external Pt100 temperature probe. The optical 
channel connectors are color coded, indicating the analyte (

oxygen

pH

  or 

temperature

),  which  is  currently  measured  and  can  be 

changed  for  each  channel.  The  air  inlet  equilibrates  the  internal 
temperature, pressure and humidity sensors with the surrounding. 
Avoid covering these holes to ensure free air circulation towards the 
internal sensors. 

 

The  micro-USB  connector  on  the  left  side  panel  provides  energy 
supply  and  data  exchange  with  the  PC.  Right-hand  side  of  it,  a 
connector  X1  for  power  and  digital  interface  (7-pins)  and  a 
connector X2 for analog output (5 pins) is located. 

 

 

 

Summary of Contents for FireSting pro

Page 1: ...1 FireSting pro FIBER OPTIC MULTI ANALYTE METER USER MANUAL ...

Page 2: ...he FireSting pro is manufactured by PyroScience GmbH Hubertusstr 35 52064 Aachen Germany Phone 49 0 241 5183 2210 Fax 49 0 241 5183 2299 Email info pyroscience com Internet www pyroscience com Registered Aachen HRB 17329 Germany ...

Page 3: ...ensor lifetime pH oxygen and temperature determination simultaneously in one sample This single device solution for multi analyte measurements has integrated atmospheric pressure and humidity sensors for precise and easy oxygen sensor calibration but also for automatic pressure compensation of the oxygen measurements Furthermore the FireSting pro offers 4 analog outputs and a broadcast mode The ne...

Page 4: ...or temperature which is currently measured and can be changed for each channel The air inlet equilibrates the internal temperature pressure and humidity sensors with the surrounding Avoid covering these holes to ensure free air circulation towards the internal sensors The micro USB connector on the left side panel provides energy supply and data exchange with the PC Right hand side of it a connect...

Page 5: ... 5 to one of the optical sensor connectors 1 to 4 for automatic temperature compensation Step 5 Prepare appropriate calibration standards as described in the respective sensor manuals see 7 Step 6 Start the logger software by clicking on the short cut Pyro Workbench on your desktop Step 7 Open the Settings Wizard by clicking on the FireSting pro picture Select the respective analyte and enter all ...

Page 6: ... a male fiber plug Remove the black caps from the plug of the sensor fiber Remove the red caps from the sensor ports at the FireSting pro the red caps should be put on again if it is not in use anymore to protect the optics Insert the male fiber plug of the sensor cable into the ST port female fiber connector of the FireSting pro and turn the bayonet coupling gently clockwise until the plug is loc...

Page 7: ...Sting pro is compatible with a broad range of optical oxygen pH and temperature sensors from PyroScience For an overview of available optical sensor types please see the PyroScience website https www pyroscience com products html ...

Page 8: ...rt Range Resolution Accuracy 1 channel for 4 wire Pt100 30 C to 150 C 0 02 C 0 5 C Internal temperature sensor Range Resolution Accuracy 40 to 125 C 0 01 C 0 3 C Internal pressure sensor Range Resolution Accuracy 300 to 1100 mbar 0 1 mbar typ 3 mbar Internal humidity sensor Range Resolution Accuracy 0 to 100 rel humidity RH 0 04 RH typ 0 2 RH Digital interface at extension port X1 7 pins UART with...

Page 9: ...tailed instructions for using the Pyro Workbench and application of optical oxygen pH and temperature sensors Manual for logger software Pyro Workbench Windows manuals for optical sensors from PyroScience oxygen pH temperature ...

Page 10: ...ecially after removal of the protective cap Prevent mechanical stress to the fragile sensing tip Prevent injuries with needle type sensors Calibration and application of the sensors is on the user s authority as well as data acquisition treatment publication The optical sensors and meter FireSting pro are not intended for medical diagnostic therapeutic or military purposes or any other safety crit...

Reviews: