13 Hardware Information
-32-
13 Hardware Information
(Changes reserved)
Build in a stable metal case with carrying handle and tough front panel.
13.1
Measuring Unit:
Sampling rate max. 100kHz
Resolution 0.01V - 0.25V, 0.005A – 0.01A (depending on sel. range)
Accuracy of the a/d converter 0.08% of FSR±1 LSB
Peak power measurement: ±5%
Reproducement: ±2%
Duration of single measurement: 20ms up to 2 seconds
Phox sensor with integrated Pt1000 sensor measures irradiance and cell temperature. Other commercially
available sensors as e.g. ISET-Sensor® can be connected using a reliable cable connection.
13.2
Pinout of sensor sockets:
Temperature (external): 4 pin female chassis socket Lumberg KFV40
Pin 1 = Current (~1 mA)
Pin 2 = Pt100 +
Pin 3 = Pt100 -
Pin 4 = Current source - (~1 mA)
Irradiance: 8 pin female chassis socket Lumberg KFV81 (plug: SV81)
Pin 1 = Irr
Pin 2 = Pt1000 (reference) +
Pin 3 = Irradiance-
Pin 4 = Current (~1 mA)
Pin 5 = Current source - (~1 mA)
Pin 6 = do not connect
Pin 7 = do not connect
Pin 8 = Pt1000 (reference) -
Measuring ranges of the PVPM:
Standard
Meas.Range
Voltage
[V]
Current
[A]
Tempe-
rature
Irradiance
PVPM2540C
25/50/100/250
5/10/20/40
-100°C -
+120°C
0 - 1300 W/m2
PVPM1000C
25/100/500/1000
2/5/10/20
PVPM1040C
25/100/500/1000
2/5/10/40
PVPM1050C
25/100/500/1000
2/5/10/50
Peak power:
up to 40kW
Ranges for current and voltage can be combined in any order
An optimal range is selected by the device automatically.
13.3
Calculation Unit:
Low power industrial PC
Flash disk 32MB (can store several 1000 results)
Watchdog, real time clock
No mechanically moved parts like hard disk, fan etc.
13.4
Display:
Bright TFT touch display with LED back-light,
resolution 480 x 272 pixels,
Operation manual for PVPM devices
PV-Engineering GmbH