REPLACING A NCP MEDIA CARTRIDGE
7
NB
to maintain optimum purity
You will then be asked to confirm if a new “RHS Pack Has Been
Fitted”, select “Yes”
You will then be asked to reset the replacement date of this pack,
selecting “Yes” will then reset the replacement date for 6 months time.
If the alarm message “Poor Outlet Quality” is ever displayed then new
packs SHOULD be fitted to both positions.
To remove a pack follow the procedures below.
•
Ensure unit is in the
“Power On”
condition and isolate the
mains water supply.
•
Open the front door cover to reveal both of the NCP media
cartridges.
•
To remove a cartridge press down on the lever on top of the
holder and gently pull away the exhausted cartridge.
(See Fig-
16)
. There may be some drops of water lost from the cartridge.
Fig-16
Fig-17
Under normal operating conditions the pack in the left hand positionwill
always exhaust first. The “Mid Qual” alarm message will indicate when
the pack is exhausted and requires replacement.
hen replacing the left hand side pack
W
remove the exhausted left hand side pack, discard it, then remove the
right-hand side pack and place it in the left hand side position. The new
pack SHOULD then be fitted to the right hand side position.
Blue release lever
White release lever
Select Neptune Manual Iss: B
Summary of Contents for SELECT NEPTUNE Analytical
Page 45: ...Select Neptune Manual Iss B...
Page 46: ...Select Neptune Manual Iss B...
Page 72: ...7 9 PREVENTATIVE MAINTENANCE AND CONSUMABLE REPLACEMENT GUIDE Select Neptune Manual Iss B...
Page 82: ...Select Neptune Manual Iss B...
Page 87: ...Select Neptune Manual Iss B...
Page 88: ...Select Neptune Manual Iss B...
Page 89: ...Select Neptune Manual Iss B...
Page 90: ...Select Neptune Manual Iss B...