3.1
Wear protective gloves and face mask! Avoid contact with the
contents of the amalgam separator!
It is not possible to open the ECO II and reuse it! Every attempt to
open the amalgam separator results in the malfunction of the device.
Moreover, the required separation rate will no longer be achieved!
The ECO II collection container must be exchanged according to the replacement intervals as defined
in the replacement interval table (cf. page 8, chapter 6).
Remove the adapter
with the tube at the
waste water inlet and
waste water outlet.
Detach the screws
from the bolts 3.1 and
take out
the full amalgam
separator by lifting it
off upwards.
Tightly close the
outlet of the full ECO
II with the sealing plug
(supplied with the
amalgam separator).
Position the full amalgam
separator horizontally
onto an even surface,
make sure that the inlet
opening is pointing
upwards!
Cut open the sanitizing
pouch (supplied with
new amalgam separator)
at one corner and pour
its content into the
inlet opening of the full
amalgam separator.
Maintenance and Replacing
the Collection Container