PureAire Monitoring Systems, Inc.
34
Push the joystick down to access the next sub menu;
Set Relay 2 Alarm Threshold,
will scroll on the digital display.
This is the gas concentration at which the instrument’s second level alarm will be
activated. To change the displayed value, push the joystick to the right to display the
second level alarm setting. The display will indicate a value between 00.0% and 25.5%.
Pushing the joystick up increases the value and pushing the joystick down decreases the
value. Press
ENTER
to accept the value.
After entering the relay state the display will default back to the
Set Relay 2 Alarm
Threshold
Menu. The display will scroll the following:
Push the joystick down to access the next sub menu;
Set Audio Alarm Threshold,
will scroll on the digital display.
This is the gas concentration at which the instrument’s audio alarm will be activated.
To change the displayed value, push the joystick to the right to display the second level
alarm setting. The display will indicate a value between 00.0% and 25.5%.
Pushing the joystick up increases the value and pushing the joystick down decreases the
value. Press
ENTER
to accept the value.
NOTE:
The O
2
system must have the audio alarm option module installed to access
this menu. If this option is installed the display will indicate N/A, (not available)
..Set Relay 2 Alarm Threshold…
20.9%
18.0%
20.9%
..Set Audio Alarm Threshold…
20.9%
19.5%
20.9%
..Set Relay 2 Alarm Threshold…
20.9%