Gebrauchsanweisung - 9/24
General information
Cleaning
Cleaning the device
– The inside of the device may not be cleaned as com-
ponents may be damaged.
– If desired, the outside of the device can be wiped with
a damp cloth (where necessary with a mild cleaning
agent). Wipe dry afterwards!
– The device is not watertight. It must not be cleaned
with a low-pressure or high-pressure cleaner.
Cleaning the lamps
The lamps only have to be cleaned if dirt gets on them
during installation or they are mistakenly touched with
bare hands! Regular cleaning is not necessary.
The lamps are highly sensitive to finger sweat and alka-
line cleaning agents!
– Wear clean and soft gloves (cotton or non-powdered
latex) to avoid applying new dirt. Only use soft, clean
cotton cloths.
– Clean with clear water (with a little washing-up liquid
when necessary).
– Wipe dry afterwards.
Cleaning the filter
(see also page 21)
The cleaning intervals are highly dependent on the oper-
ating conditions.
– Check the filter on the inside and outside as well as the
protective grille behind the filters for visible soiling on a
regular basis.
Dry cleaning of the filter and protective grille: Vacuum
cleaner (depending on the degree of soiling).
Moist cleaning (filter only!): water and detergent, also a
dishwasher.
Maintenance
Maintenance must be carried out after 9000 hours or 1
year.
The next maintenance appointment is indicated on the
maintenance sticker as well as the service booklet.
Please make an appointment in good time with Customer
Service (see page 2). The maintenance is carried out by
replacing the reflector and lamps on-location.
Assessment of the condition after 10 years
When the device reaches an age of 10 years after the
date of manufacture, the condition of the device must be
assessed by a certified service technician from JK-Inter-
national GmbH, JK-Global Service Department before the
device may be used further.
If this device assessment is not carried out, personal inju-
ry or material damage may occur.
Any liability is excluded if the device remains in use 10
years after the date of manufacture without its condition
being assessed.