6.3 Temperature
Factory Preset
Unit F
Analog Output
4-20mA
Analog High
200
Analog Low
0
Thermocouple J
6.3.1 Setup Temperature Sensor Type
1. Press and hold the “E” and “M”
buttons for three seconds. The
screen will display “cond”.
2. Press the “M” button until the
display shows “rAnG” and then a
sensor type.
3. Press the “UP” arrow to change to
“J” if it is not showing “J”.
4. Press the “M” button to go to the
next parameter or press the “E”
button to exit the setup.
6.3.2 Setup Temperature Unit
1. Press and hold the “E” and the “M”
button for three seconds and the
screen will display “cond”.
2.
Press the “M” button until the
display shows “Unit” and then “c” or
“F”.
3. Press the “UP” button to change to
unit.
4. Press the “M” button to go to the
next parameter or press the “E”
button to exit the setup.
6.3.3 Setup Analog Output Mode
1. Press and hold the “E” and “M”
buttons for three seconds. The
screen will display “cond”.
2. Press the “M” button until the
display shows “A.OUT”.
3. Press the “UP” arrow to change the
analog output mode: 0-1V, 0-10V,
1-5V, or 4-20mA.
4. Press the “M” button to go to the
next parameter or press the “E”
button to exit the setup.