Instructions to User 

This Fingertip Pulse Oximeter is a kind of innovated medical device with non-invasive and continous 
features for artery SpO2 and PR detection. Being portable, it is able to measure SpO2 and PR values 
quickly and precisely. This product is medical device, which can be used repeatedly. 

 

General Description 

Haemoglobin Saturation is the percentage between the capacity of Oxyhemoglobin (HbO2) that 
compounded with oxygen and that of all combinativable haemoglobin (Hb) obin (HbO2) in blood. In 
other words it is the saturation of Oxyhemoglobin in blood. It is a very important physiological 
parameter for Respiratory and Circulation Systems. Many respiratory diseases could reduce 
haemoglobin saturaion in human blood. Moreover, factors such as Automatic Organic Regulation 
Malfunction caused by anaesthesia, trauma resulted from major operation and some medical 
examination can also cause problems in oxygen supply, which might reduce human haemoglobin 
saturation. As a result, such symptoms as megrim, vomiting and asthenia might appear to patients. 
Hence, it is very important to know haemoglobin saturation of patient timely in clinical medical aspects. 
The Fingertip pulse oximeter features in small volume, low power consumption, convenient operation 
and portability. It is only necessary for patient to put one of finger into fingertip photoelectric sensor for 
diagnosis, and the display screen will directly show the measured value of hemoglobin saturation. It has 
been proved in clinical experiments that it possesses rather high precision and repeatability. 

Measurement principle 

The principle of the oximeter is as follows: An experience formula of data process is established by 
exerting Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive Hemoglobin 
(R Hb) and Oxyhemoglobin (O2 Hb) in glow and near-infrared zones. Operation principle of the 
instrument is to combine Photoelectric Oxyhemoglobin Inspection Technology with Capacity Pulse 
Scanning and Recording Technology, so that two lights with different wavelength (660nm glow and 
905nmnear-infrared light) can be focused onto human nail through perspective clamp finger-type 
sensor. Then measured signal can be obtained by a photosensitive element, information acquired 
through which will be shown on two groups of LEDs through 
process in electronic circuits and microprocessor. 

Diagram of Operating Principle 

1.

 

Infrared-ray receiving tube 

2.

 

Infrared-ray transmitting tube 

Precautions for Use 

1.

 

Do not use the pulse oximeter together with MRI or CT equipment. 

2.

 

Explosion hazard: Do not use the pulse oximeter in an explosive atmosphere.  

3.

 

The pulse oximeter is intended only as an adjunct in patient assessment. Doctors should make the 
diagnosis in conjunction with clinical manifestation and symptoms. 

4.

 

Check the pulse oximeter sensor application site frequently to make sure that the circulation and 
skin integrality of patient are under good condition. 

5.

 

Do not stretch the adhesive tape while applying the pulse oximeter sensor. This may cause 
inaccurate reading or skin blisters.  

6.

 

Please read the manual carefully before operation. 

7.

 

Prolonged use or the patient’s condition may require changing the sensor site

 periodically. Change 

sensor site and check skin integrality, circulatory status and correct alignment at least every 2 hours. 

8.

 

Inaccurate measurements may be caused by autoclaving, ethylene oxide sterilizing, or immersing 
the sensors in liquid. 

9.

 

Significant levels of dysfunctional hemoglobins (such as carboxyl hemoglobin or methemoglobin) 
may cause inaccurate reading. 

10.

 

Intravascular dyes such as indocyanine green or methylene blue may cause inaccurate reading. 

Summary of Contents for PO-210B

Page 1: ...OX PO 210B BEDIENUNGSANLEITUNG Pulse Oximeter PO 210B Instruction manual Pulsiox metro PO 210B Manual de instrucciones Oxym tre de Pouls PO 210B Mode d emploi Pulsossimetro PO 210B Instruzioni per l u...

Page 2: ...feld von gr ter Wichtigkeit Das Pulsoximeter ist klein und tragbar hat einen geringen Stromverbrauch und eine komfortable Bedienung Messprinzip Das Prinzip des Pulsoximeters Es wird eine mathematische...

Page 3: ...n Verordnungen und Recyclinghinweise zur Entsorgung bzw zum Recycling des Ger tes und der Ger tekomponenten Anwendungsbereich Verwendungszweck Das Pulsoximeter ist f r Kinder unter 12 Jahren geeignet...

Page 4: ...n neben Restore angezeigt wird dr cken Sie die Taste lange 3s und alle Einstellungen werden wieder auf die Werkseinstellungen zur ckgesetzt Achtung Bei Verwendung des Ger ts im Freien oder unter stark...

Page 5: ...her Schock Grad Typ BF Wartung und Instandhaltung 1 Laden Sie den Akku rechtzeitig auf wenn der Akkustand zu niedrig ist 2 Reinigen Sie die Oberfl che des Oximeters bevor Sie es zur Messung am Patient...

Page 6: ...Laden Sie den Akku auf Garantie und Service Wir leisten 1 Jahre Garantie f r Material und Fabrikationsfehler des Produktes Die Garantie gilt nicht im Falle von Sch den die auf unsachgem er Bedienung b...

Page 7: ...Vor direkter Sonne sch tzen Die CE Kennzeichnung bescheinigt die Konformit t mit den grundlegenden Anforderungen der Richtlinie 93 42 EEC f r Medizinprodukte Ein Aus Technische nderungen vorbehalten P...

Page 8: ...s is established by exerting Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive Hemoglobin R Hb and Oxyhemoglobin O2 Hb in glow and near infrared zones Operation principle...

Page 9: ...and SpO2 sensor Features OLED Display Product adopts double color OLED display can show the six different display mode Low power consumption continuously work for more than six hours with one recharg...

Page 10: ...item then press button 3s to setup data Choose or to plus or minus values Menu 1 Menu 2 Detailed Description of Product Functions 1 Display type OLED Display 2 SpO2 Measurement range 70 99 Accuracy 2...

Page 11: ...er must use it in the following environments Radiation Test Compliance Electromagnetic environment guidance RF interference CISPR 11 Group 1 RF signal of Pulse Oximeter is simply created by its intern...

Page 12: ...d part European Representative WEEE 2002 96 EC Serial number Refer to instruction manual Lot number Caution Date of manufacture IP22 International Protection degree Manufacturee Humidity range Tempera...

Page 13: ...medici n del pulsiox metro funciona de la siguiente manera Se establece una f rmula de experiencia de procesamiento de datos aplicando la Ley de Lambert Beer seg n las caracter sticas de absorci n del...

Page 14: ...utilizar para medir la saturaci n y la frecuencia card aca de los humanos a trav s del dedo Se puede utilizar en hospitales familias escuelas y centros m dicos NOTA 1 Es posible que detecte que las im...

Page 15: ...zonas con mucha luz ajuste el brillo de la pantalla adecuadamente Seleccione un menor brillo para ahorrar bater a 2 Men 2 Cuando aparezca el indicador en el men Sounds Setup mantenga pulsado el bot n...

Page 16: ...da arlo Accesorios del producto 1 Manual de usuario 2 Cable de carga 3 Adaptador opcional Gu a y declaraci n de fabricaci n Radiaci n electromagn tica para otros equipos y sistemas El pulsiox metro es...

Page 17: ...u s de 8 segundos 2 El dispositivo tiene poca bater a 1 Normal 2 Cargue la bater a S mbolos y Definiciones Symbol Description Symbol Description Protecci n y aislamiento del dispositivo el ctrico Repr...

Page 18: ...n et r p tabilit Principe de mesure Le principe de mesure de l oxym tre fonctionne de la mani re suivant il tablit une formule d exp rience de traitement de donn es en appliquant la loi de Lambert Bee...

Page 19: ...on Pr vue L oxym tre de doigt peut tre utilis pour mesurer la saturation et la fr quence cardiaque des humains travers le doigt Il peut tre utilis dans les h pitaux en famille dan les coles ou les cen...

Page 20: ...nt plus de trois secondes tous les r glages retourneront la configuration d usine Attention Si le dispositif va tre utilis l ext rieur ou dans des zones avec beaucoup de lumi re ajustez la brillance d...

Page 21: ...hargez la batterie que lorsque le niveau de la batterie est faible 2 Nettoyez la surface du bout du doigt avant de l utiliser 3 Conservation du produit 10 C 40 C 14 104 F et humidit de 10 80 4 Il est...

Page 22: ...2 L oxym tre pourrait tre endommag 1 Chargez la batterie 2 Veuillez contacter votre distributeur Les indicateurs s teignent subitement 1 Le dispositif s teint automatiquement s il ne d tecte pas de s...

Page 23: ...segue Viene stabilita una formula matematica che si avvale della Legge di Lambert Beer in conformit con le caratteristiche dello spettro di assorbimento dell emoglobina riduttiva R Hb e dell ossiemogl...

Page 24: ...er misurare la saturazione e la frequenza cardiaca nelle persone e si pu utilizzare in ospedali casa scuole e centri medici NOTA 1 possibile che le immagini presenti in queste istruzioni siano differe...

Page 25: ...lta luce regolare la luminosit del display di conseguenza 2 Men 2 quando appare l indicatore sul men Sounds Setup premere il pulsante e tenerlo premuto per oltre tre secondi per accedere al secondo me...

Page 26: ...i fabricazione radiazione elettromagnetica per altri apparecchi e sistemi Questo pulsossimetro progettato per essere utilizzato in un contesto elettromagnetico specifico Gli utenti devono utilizzarlo...

Page 27: ...2 Ricaricare la batteria Simboli e Definizioni Simbolo Descrizione Simbolo Descrizione Protezione e isolamento del dispositivo elettrico Rappresentante dell Unione Europea Raccolta differenziata WEEE...

Page 28: ...ne question s il vous pla t contacter Se si dispone di una domanda si prega di contattare Novidion GmbH Fuggerstr 30 51149 K ln Tel 49 2203 9885 200 Fax 49 2203 9885 206 info novidion de www pulox de...

Reviews: