Pulse Oximeter tester) could measure the available pulse wave with the amplitude of 6% of the
simulation pulse wave amplitude.
11.
Interference Resistance Capacity against ambient light: Device works normally when BIO-TEK INDEX
Oximeter tester exerts interfering signal test.
Classification
1.
Management Class for Medical Devices: II Equipment
2.
Anti-electric Shock Type: Internally powered equipment
3.
Anti-electric Shock Degree: Type BF equipment
Maintenance and Preservation
1.
Charge batteries timely when the battery power is too low.
2.
Clean the surface of fingertip oximeter before it is used to diagnose patients.
3.
It is recommended to preserve the product in -10 - 40°C (14 - 104°F) and humidity is 10%-80%.
4.
It is recommended that the product should be kept dry any time. A wet ambience might affect its
lifetime and even damage the product.
Product Accessories
1.
User manual
2.
Charging cable
3.
Adapter (optional)
Guidance and Manufacturer’s Declaration
- Electromagnetic Radiation for Other Equipments
and Systems
The Pulse Oximeter is designed to be used in specified electromagnetic environment. Users of the
Pulse Oximeter must use it in the following environments.
Radiation Test
Compliance
Electromagnetic environment
guidance
RF-interference CISPR 11
Group 1
RF signal of Pulse Oximeter is
simply created by its internal
function. Therefore, its RF
interference is very low and is
not likely to cause any
interference to nearby
electronic equipment.
RF-interference CISPR 11
Class B
The Pulse Oximeter applies to
all establishments, including
domestic establishments and
those directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes.
Troubleshooting
Problem
Possible Reason
Solution
The SpO2 and Pulse Rate can
not be shown normally
1. Finger is not properly
positioned
2. P
atient’s SpO2 is too low to
be measured.
1. Place the finger properly and
try again.
2. Try again; If you are sure the
device works all right, go to a
hospital timely for exact
diagnosis.
The SpO2 and Pulse Rate are
shown unsteadily
1. Finger is not placed inside
deep enough.
2. Finger is trembling or the
patient is moving.
1. Place the finger properly and
try again.
2. Please remain at rest.
The device can not be turned
on
1. Inadequate power or power
off.
2. Oximeter might be damaged.
1. Charge battery timely.
3. Please contact the local
customer service.
Display is suddenly off
1. The device automatically
1. Normal.