Latex Free

MaxShield™ Mask with

Anti-Asphyxia 

With Leak Elbow 

22mm Female

Interchangeable Elbow

Compatible

BiTrac Select

Intended Use 

The BiTrac MaxShield™ mask is an accessory for use with CPAP or Bilevel systems 

for ventilatory support. A minimum 

of >3.0cm of H

2O

 at the mask is required.

Contraindications

1)  This device should not be used on uncooperative or unresponsive patients.

2)  This device may not be suitable for use on patients unable to remove the mask or are taking medication which may  

  cause vomiting, have nocturnal vomiting, hiatal hernia or impaired cardiac sphincter function.

3)  This mask may not be suitable for use on patients with the following conditions: glaucoma, recent eye surgery or dry  

  eyes.  

Warnings

1)  This mask has an integral leak port that provides a continuous leak path in the patient circuit. It is recommended 

  that the circuit have an additional exhalation device. You may need to adjust the pressure level and oxygen therapy 

  to compensate for the additional leak of the exhalation device.

2)  This device is not suitable for providing life support ventilation.

3)  Any unusual chest discomfort, shortness of breath, stomach distension, belching, severe headache, drying eyes, eye  

  pain, eye infections or blurred vision when receiving positive airway pressure or upon awakening should be reported to  

  the clinician. 

4)  To minimize risks of vomiting during sleep, avoid eating or drinking for three (3) hours before use.

5)  Make sure the mask and headgear are the correct size for the patient’s face. If excessive discomfort is experienced,  

  adjust the mask and head gear. If the discomfort continues, consider an alternative mask.

6)  This CPAP mask should only be used with CPAP systems recommended by your physician or respiratory therapist. A 

  mask should not be used unless the CPAP system is turned ON and operating properly. The vent hole or holes    

  associated with the mask should never be blocked. 
 

Explanation of the Warning: CPAP systems are intended to be used with special masks with connectors which    

  have vent holes to allow continuous flow of air out of the mask. When the CPAP machine is turned on and functioning  

  properly, new air from the CPAP machine flushes the exhaled air out through the attached mask exhalation port.   

  However, when the CPAP machine is not operating, enough fresh air will not be provided through the mask, and   

  exhaled air may be rebreathed. Rebreathing of exhaled air for longer than several minutes can in some circumstances  

  lead to suffocation. This warning applies to most models of CPAP systems.

Before Use

Verify entrainment valve function on anti-asphyxia elbow:

•  With the unit turned OFF, the entrainment valve flap should 

  be open, allowing user to breathe room air.

•  Turn CPAP or Bilevel unit ON.

•  The flap valve should close, allowing the user to receive airflow from the unit.

•  If the valve does not function in this manner, replace the mask.

•  Do not block the opening on the entrainment valve.

•  Check valve before each use for malfunction or blockage caused by patient secretions. 

Use of Interchangeable Elbow Feature: 

1.  Remove the existing elbow from the mask by holding the adapter ring and twisting the elbow counter clockwise    

 

to disengage the bayonet lock (Refer to Figure B.)

2.  Gently pull the elbow from the mask. (Refer to Figure C.)

3.  Gently press the desired Interchangeable Compatible Elbow into the mask adapter ring and turn clockwise until    

 

fully seated. Do not use excessive force. 

Bayonet Lock References:

Operating Specifications:

 5°C to 40°C 

at humidity range of 15% to 95%

Storage Specifications:

 -20°C to 

60°C at reltive humidity up to 95% non-
condensing

Made in USA

M

Pulmodyne Inc.

2055 Executive Drive
Indianapolis, IN 46241 USA

www.pulmodyne.com

P

Caution:

  Federal Law (USA) restricts this device to sale    

by or on the order of a licensed physician.

Single Patient Use

Pulmodyne is a registered trademark of 
Pulmodyne, Inc. BiTrac MaxShield 
is a registered trademark of Pulmodyne, 
Inc.

2434D-08-11-2020

C

2797

1

Figure A.

Figure B.

Figure C.

EN

-Medical Device

Summary of Contents for BiTrac Select

Page 1: ...r be blocked Explanation of the Warning CPAP systems are intended to be used with special masks with connectors which have vent holes to allow continuous flow of air out of the mask When the CPAP mach...

Page 2: ...ei Patienten verwendet werden die unkooperativ sind oder keine Reaktion zeigen 2 Dieses Ger t ist evtl nicht geeignet f r Patienten die nicht in der Lage sind die Maske abzunehmen oder die eine Medika...

Page 3: ...e muni d un raccord femelle de 22 mm Raccord coud interchangeable compatible BiTrac Select Utilisation pr vue Le masque BiTrac MaxShield est un accessoire pr vu pour les syst mes CPAP ou Bi Level dans...

Page 4: ...litys CPAP j rjestelm t on tarkoitettu k ytett v ksi sellaisten erikoismaskien kanssa joiden liittimiss on erityiset ilma aukot jatkuvan ilmavirtauksen sallimiseksi maskista ulos Kun CPAP laite kytket...

Page 5: ...mpatibile nell anello adattatore della maschera e ruotare in senso orario finch completamente alloggiato Non esercitare una forza eccessiva Specifiche di funzionamento 5 C 40 C nell intervallo di umid...

Page 6: ...verkocht Sem l tex M scara MaxShield com conector f mea de 22 mm anti asfixia com cotovelo de fuga Compat vel com o cotovelo intercambi vel BiTrac Select Utiliza o prevista A M scara BiTrac MaxShield...

Page 7: ...a Wyprodukowano w USA Przestroga Wed ug przepis w prawa federalnego USA urz dzenia te mog by sprzedawane wy cznie lekarzowi lub na jego zlecenie Latexfri MaxShield mask med anti asfyxihus med l ckagek...

Page 8: ...kullan lmal d r Maske CPAP sistemi A IK de ilse ve d zg n al m yorsa kullan lmamal d r Hava alma deli i ya da delikleri hi bir zaman bloke edilmemelidir Uyar n n A klamas CPAP sistemleri maskeden s re...

Reviews: