ADG400 User Manual
14
99998 Issue 1, November 2022
OPERATION
3.6.6.1 Transition Delay
When changing from full dry (Dryer mode) to a 2-stage set point, there is a 2-minute
delay before the MFC5 wet flow set point is selected to allow the upstream pipework to
be purged. This is reflected by the target/set point of MFC5 on the status screen being
set to 0, and the TD indicator being illuminated.
3.6.7 Settings Menu
Item
Description
Reset Calibration
Restore either a single point or all of the
Standard
calibration
table from the factory settings stored in flash memory.
Edit Tables
Launch either the Set Point Editor or Profile Editor.
Saturator Filling
Launch the guided saturator filling mode.
Screensaver
Time before enabling the screensaver.
Protection
Protect Settings menu or Main screen from tampering with a
simple fixed passcode.
Back
Return to Main screen.
Figure 11
Settings Screen