RT9200-01 Rev B Page 2
3.
Place the sump on the floor and stack the pump on top of the sump. Place the body model on a flat tabletop or
similar surface nearby the pump and sump.
Note:
The STAAR has a simulated circulatory system with pressure presets that are designed with the intention that
the sump and pump are to be placed on the floor and the simulator body is to be placed at table height. If the pump
and sump are placed even with the body, the preset pressures will not be accurate.
4.
Connect the tubing to the pump, sump and body. There are three tubes with red, green and yellow labels.
Connect the corresponding colors of plugs to sockets, ensuring that metal plugs fit into metal sockets and plastic
plugs fit into plastic sockets.