RT9200-01 Rev B Page 8
C.
Disassembly of the STAAR
1.
To disassemble the STAAR, begin by turning off the pump.
2.
To drain water out of the model, open a pathway in the tubing by either:
a.
Opening
the introducer sheath’s sidearm stopcock.
b.
Disconnecting the blood pressure monitoring extension line while the ER-REBOA catheter remains in the
model.
3.
If you are using the vena cava, open the stopcock on the pressure infusion bag and allow adequate time for the
bag to deflate. Disconnect the blue vena cava tube from the extension line.
Note:
The water in the vena cava will not have a reservoir to easily drain into, be prepared for a small amount
of water come out of the tube.
4.
Tilt the model to facilitate drainage. Hold until as much of the water has left the model as possible. When the
model is drained, unplug the green “MODEL IN” and red “MODEL OUT”
tubes and connect them to each other.
5.
Turn on the pump circulatory flow
and while the pump is running, disconnect the “SUMP OUT” yellow tube from
the sump.
Note
: This will allow the pump to run dry while pumping water into the sump. Allow the pump to
run dry for approximately 10-
30 seconds. Turn off the pump and disconnect the green “MODEL OUT” tubing
from the pump. Open the sump and drain the remaining water in the tubing into the sump.
6.
Empty the sump and return the algaecide solution bottle into the sump case.
7.
Disconnect the power supply, pressure monitoring cable and USB cable from the pump.
8.
Unplug the GFCI outlet protector, the extension cord and power supply.
9.
Replace the tubing into the body model’s cavity.
10.
Disconnect the USB cable from the tablet and power down the tablet.
11.
Replace all components, support equipment, accessories and/or kits into the transport case and to their
respective places.
12.
Ensure the case can close and be zipped up for storage.
Note
: There is a second zipper that allows the case to expand if needed. If desired, press the zippers into the
securing mechanism on the front of the case.