36
Picture 48: Device information
Sensor Information
The
settings
mode displays following information relevant to the sensor:
Software Version
Serial number
Internal Temperature (temp. inside the sensor)
Information is read out any time you click on
Connect
.
15.2 Basic configuration
Average
In this setting to be averaged single values are determined for a measurement output.
Values from 1 to 50 can be
entered here.
Example 1:
Measuring frequency: 20
Hz Average value: 1
Output frequency: 20 Hz
Example 2:
Measuring frequency: 20
Hz Average value: 10
Output frequency: 2 Hz
Measurement frequency
This setting determines the measuring frequency of the LM_100. A 0 is entered for an automatic
measuring frequency.
In
the
continuous
measurement mode, values
from 0 to 20 Hz can be set
here.
In
Continuous
Tracking
mode, the following values
can be set;
0, or 20 to 100
In the external trigger mode,
values from 0 to 100 can be
entered.
Picture 49: Average
Picutre 50: Measurement frequency