HOW IT WORKS
REV G /2021
P5
Clean Time
: The period of time the device is producing UV-C.
No Motion Time
: The length of time the device will allow to pass before producing UV-C after the motion sensor
has communicated an absence of movement. Suggested for high tra
ffi
c areas, mobile carts, and personal desks
is 240 or 300 seconds.
Wait Time
: Downtime scheduled between cleaning cycles that are unrelated to motion sensor activity.
After changes are made (1) you must mouse over and click the
Save to Device
button (2).
Data Retrieval and Making Changes
Set the Device to your local time by mousing over and clicking the
Set Device To Local Date & Time
button
located on the
Home
tab.
Date & Time
(1)
(2)
Required:
Plug in micro USB data cable. For data transfer to PC.
Downloaded software from the product website is also required.
Follow direction below if data retrieval from the device is needed or changes to the
device pre-set times is needed. Product is set to factory settings. Factory settings are
for low tra
ffi
c areas and disinfection is accomplished via cumulative cycles when the user
has stepped away from the unit. If the product is used in high tra
ffi
c area or you notice
the unit is not cycling as often as you would like you can change these settings or contact
your reseller or Proximity for recommended settings based on your application.