4.3

 

Tab 1: “Profiles” 

 

The sens07 app can manage up to 3 very different profiles defining 
the way how sen07vest is operating. 
This screen shows the 3 profiles including profile number (1 to 3), 
title and short description. 
The button “Reorder” on the upper right corner adds a “handler” to 
each profile, allowing to move the profile up or down. 

Clicking on the profile name goes to the profile configurator. 
The last segment shows the actual 

profile switch mode

 defining 

profile switching when holding the self-test button. 
The button “start transfer” launches the process to transfer the 
profiles to the sens07 inflator. 

 

 

User Manual 

sens07vest 

May 2020 

 

Page 19 

4.3.1

 

Profiles configurator (basic) 

The drowning algorithm is based on 3 depth zones: 

Safe zone

Diving zone

 and 

Inflation zone

A simple inflation countdown starts at the 
specified value when entering into the 

Div-

ing Zone

. Inflation is triggered when infla-

tion countdown reaches zero. 
This inflation countdown freezes when re-
turning into the 

Safe Zone

. Staying in the 

Safe Zone

 for the specified time resets the 

countdown to the initial start value. 
Entering into the 

Inflation zone

 will immedi-

ately inflate. 
Pressing the 

Suspend Button

 under water 

will freeze the countdown for the specified 
time (Inflation Delay). This feature is op-
tional and is activated by the slider “

Enable 

Suspend Button

”. 

Combat diver can disable the 

LED Signal

 un-

der water for this specific profile. 
The last segment 

Profile Info

 offers to 

name the profile and to add any kind of free 
description for better understanding. 
Save changes before returning to the main 
menu. 

 

 

Summary of Contents for sens07vest

Page 1: ...User Manual sens07vest May 2020 Page 1 User Manual Mai 2020...

Page 2: ...Inflation reason report 15 3 2 2 Red LED Battery low 15 4 sens07 app 16 4 1 Introduction 16 4 2 Installation and registration 17 4 3 Tab 1 Profiles 18 4 3 1 Profiles configurator basic 19 4 3 2 Profil...

Page 3: ...r after the dive in order to prevent inflation of the bladder The sens07vest is designed to protect free divers and spear fisher up to 50 meters under water when using a 95gr CO2 cylinder Be low this...

Page 4: ...nserted before each dive User Manual sens07vest May 2020 Page 7 1 4 Activation on water contact The sens07 inflator immediately starts its auto supervision when wa ter can be sensed by the integrated...

Page 5: ...p User Manual sens07vest May 2020 Page 9 1 8 Customization using the sens07 app The inflation parameters and features called profile can be customized using the sens07 app and are wirelessly transferr...

Page 6: ...ns07vest May 2020 Page 11 2 2 Suspend upcoming inflation while diving Pressing the self test button when the sens07vest is submerged will suspend freeze the risk inflation countdown for the specified...

Page 7: ...Perform the self test as described after replacing the actuator module User Manual sens07vest May 2020 Page 13 3 LED signals The inflator head has 2 front LED for important information and 3 back LED...

Page 8: ...self test button Blackout warning Inflation occurred sens07 inflator is functional Timed inflation active Timed inflation standby User Manual sens07vest May 2020 Page 15 3 2 System LEDs on the back s...

Page 9: ...charge available for IOS and for Android devices and can be downloaded and installed in the usual way from the Apple App store or Android Play store A registration code for private or cooperate users...

Page 10: ...basic The drowning algorithm is based on 3 depth zones Safe zone Diving zone and Inflation zone A simple inflation countdown starts at the specified value when entering into the Div ing Zone Inflatio...

Page 11: ...eing at least 10 seconds in the Diving Zone Upcoming inflation can only be stopped by pressing the button Combat diver can disable the LED Signal un der water for this specific profile The last segmen...

Page 12: ...support if activated 3 After transfer check for the blinking green LED confirming suc cessful data transfer 4 Confirm seeing the green LEDs by clicking Yes on the app 5 Click on Retry in case of no b...

Page 13: ...d seconds until inflation There are 3 sliders for additional options The countdown start after transfer or on first water contact Countdown aborts when sens07 is taken out of the water Countdown abort...

Page 14: ...Bladder Verify the bladder for any hole or other kind of dam age It is recommended to inflate the bladder every 2 years to check it for holes Straps Verify if the straps are well sewed and the buckle...

Page 15: ...e that the manual cord is accessible from outside and that the cord runs directly without any diversion to the sens07 base The cord is equipped with a Velcro tag near to the red handle This Velcro tag...

Page 16: ...ne starts IC Remaining in safe zone resets IC Entering diving zone starts IC Expired IC inflates sens07vest 0 min 1 min Basic Configuration Mode Safe zone 3 m Inflation countdown 2 min Diving zone 20...

Page 17: ...m Blackout countdown 15 sec Diving zone 20 m 15 m Warning zone Blackout 15 sec User Manual sens07vest May 2020 Page 33 7 Advanced explanations trouble shooting 7 1 Unsuccessful optical data transfer A...

Page 18: ...uration of the inflator head can be tested without the vest CO2 cylinder base unit and actuator module Just immerse the inflator head into water and observe the LED for water zone indicator and inflat...

Reviews: