![Protec FI-75 MAMMO Operation Manual Download Page 24](http://html.mh-extra.com/html/protec/fi-75-mammo/fi-75-mammo_operation-manual_1619430024.webp)
CP - OPTI 2
3
Trouble Shooting
English
2
General
2.1 Unit does not switch on
•
Ensure that electrical socket has power supply.
•
Check machine fuse in main switch.
•
While power switch is on, check the following components: Voltage on contact
of main switch - if no voltage: - change main switch. Check input voltage at
electronics. If the voltage is normal, exchange the electronics. if no voltage:
check the cable wiring harness.
3
Drive
3.1 Film transport does not run, fan runs
•
When placing processor cover on, the cover switch should be activated, re-
adjust if necessary.
•
Cover switch has no current passage when activated: Replace.
•
Check screwing of chain wheel on motor- and driveshaft.
3.2 Machine does not start automatically
•
Film switch is not correctly positioned or operator wire is bent. Re-adjust film
switch and operator wire.
•
Check following parts: Film switch, wiring of film switch and electronic.
•
Check wiring from electronic to the connections of components (motor, fan,
dryer heating, solenoid valve). If the connections have no fault then electronic
is possibly defective.
3.3 Machine doesn't stop automatically
•
Display „Infeeding film“ is permanently illuminated: Wire band of film switch is
jammed. Readjust wire.
•
Check following parts: Film switch, wiring of film switch and electronic.
3.4 Drive motor does not run
•
Check cover switch.
•
Check drive motor: If voltage can be detected on motor, then motor is defec-
tive.
•
Dryer fan runs but no voltage on motor: Interruption in the wiring.
3.5 Machine stops before film comes out
The cycle time is the processing time which runs off after a film has passed the film
switch. Activate the switch in the infeed tray with a film and remove the film. Meas-
ure the time until the processor stops automatically.
Following times can be measured (+/- 5%):
Change cycle time if necessary. This can be accomplished by changing the position
of the jumper on the upper side of the electronics.
Pleasse use as replacement fuses only the PROTEC
®s.
These
fuses are optimized for use under existing conditions.
Standard machine:
125 s
Jumper 2-3
Mammography:
155 s
Jumper 1-2
Optional type:
190 s
Jumper removed