21
CP - OPTI
Installation Data
Installation Data
75cm (29.5’‘)
7
0
c
m
(
2
7
.6
’‘)
4
2
c
m
(
1
6
.5
’‘)
77cm (30.3’‘)
59cm (23.2’‘)
59cm (23.2’‘)
Optional worktable or
closed cabinet
(accessories)
1
2
4 5
7
3
3
0
c
m
(1
2
’‘)
1
3
0
c
m
(5
1
’‘)
8
5
c
m
(3
4
’‘)
2
0
c
m
(8
’‘)
50cm (20’‘)
60x40cm (24x16’‘)
6
8
1.
Wall socket
220-240 V, 16 A resp. 110 V,
15 A (depending on machine
model). Power lead should be
equipped with Earth-Leakage
Switch, 25 A / 30 mA nominal
error-current. In addition, a
power control switch can be
installed.
2.
Fresh water connection 3/4”
with stop cock, permissible
pressure 2-10 bar, water tem-
perature 5-25 °C.
3.
Wasserabfluß PVC-Rohr
∅
50
mm (2“) mit Siphon.
4.
Drainage resp. collecting con-
tainers for used developer.
5.
Drainage resp. collecting con-
tainers for used fixer.
6.
Storing space for replenish-
ment tanks: Below machine or
externally.
7.
Ventilation of darkroom is nec-
essary.
8.
Sink with freshwater and flexi-
ble hose. Inner dimensions
minimum (LxWxH)
60x40x30 cm (24x16x12“).
Sink
Darkroom
Measures and positions are recommendations