Main troubleshooting

Malfunction

Reasons

Solution

Low vacuum degree

1. Insufficient vacuum pump oil.
2. Pump oil emulsion, dirty
3. Pump oil inlet plugged.
4. Leakage in pump connection.
5. Components worn out.

1. Add oil to central line
2. Put new oil
3. Clean oil inlet.
4. Check connection
5. Maintain the machine, especially o-ring,
washer and other sealing parts.

Vacuum pump inject
oil.

1. Excessive oil volume.
2. Entrance pressure too high.

1.Discharge some oil to proper level
2.Run Recovery function first.

No display

1. Fused (in Power cable connection
box, or PCA)
2. PCA burnt.
3. Power cable loosened.
4. LCD not work

1. Change fuses.
2. Change PCA.
3. Connect power cable reliably.
4. Change LCD.

Recovery does not
stop

1. Leakage in automotive A/C or

equipment pipeline.

2. Compressor not work
3.Pressure sensor does not work

Remarks: In winter, it is normal
that recovery takes longer time.

1. Make leakage test. Machine leakage test

with referenc to service manual.

2. Change compressor.
3. Fasten pressure sensor connection to

PCA, or change the pressure sensor.

No change in
recovery volume

1. No refrigerant in A/C.
2. Support screw of gas cylinder load
cell not loosened.
3.Gas cylinder load cell not work or
PCA failure

1. Stop recovery.
2. Unsrew the protection screw, as chapter

“Operation preparations”.

3. Calibrate gas cylinder load cell, or change

the load cell, or change PCA.

While auto A/C has
refrigerant,
equipment displays
alarm 005

Low pressure switch plug

disconnected from PCA socket.

Fasten low pressure switch plug.

High pressure alarm
004 but gas cylinder
gauge does not show
excessive pressure
value

1. High pressure switch plug
disconnected from PCA socket.
2. Pipeline connecting compressor
exit blocked.

1. Fasten high pressure switch plug.
2. Inspect the hoses and connections between
compressor exit and tank blue hand vale.

No charge or slow
charge.

1.Insufficient refrigerant in
equipment
2.Charge line problem.

1. Fill equipment tank with more refrigerant.
2..Check charge line, including tank red

valve, tank red hose, solenoid #5, solenoid
#9 (high side), solenoid #11 (long side),
service hoses and HP/LP quick couplers.

Summary of Contents for AC-200

Page 1: ...PROTEC AC 200 USER MANUAL AC 200 USER MANUAL...

Page 2: ...he storage cylinder in this unit contains liquid refrigerant Overfilling of the cylinder can cause violent explosion Do not disable the overfill safety feature Always keep the cylinder on the load cel...

Page 3: ...FUNCTION TABLE 3 PART DESCRIPTION 4 OPERATION PREPARATIONS 5 EQUIPMENT CONNECTION Recovery vacuum oil injection charge auto mode 6 RECOVERY 7 VACUUM 7 OIL INJECTION 7 CHARGE 8 AUTO MODE 9 SYSTEM SETT...

Page 4: ...er load cell 10g New oil bottle capacity 250ml Used oil bottle capacity 400ml Max Pressure 20bar Charge speed 2Kg Min max LCD display 4 3 inch TFT full color High pressure gauge range 1bar 40bar Low p...

Page 5: ...rigerant from equipment gas cylinder to automotive A C system Auto mode Performs the selected functions in a fully automatic sequence The machine will stop automatically once all the selected function...

Page 6: ...erant gas cylinder load cell 4 Drier filter 5 Manifold assembly 6 Assembly of solenoid valve and check valve 7 Base 8 Condenser and cooling fan 9 Compressor 10 Back cover1 11 Front wheel 12 Rear wheel...

Page 7: ...on the machine select corresponding refrigerant type run recovery to fill equipment with refrigerant After one refrigerant tank is filled fill other tank with its corresponding refrigerant It is recom...

Page 8: ...ion For recovery vacuum oil injection charge Auto Mode Warning Except the situations clearly stated in the manual during recovery and all other equipment operations please maintain the vehicle engine...

Page 9: ...sture oil and foreign particles are separated from the refrigerant before it is stored in the internal refrigerant cylinder Vacuum Select Vacuum icon and press ENTER to start the process The Vacuum pr...

Page 10: ...ual oil injection is effected by pressing right arrow key on control panel Please be aware that there is no oil load cell equipped in the machine the oil injection volume is calculated by injection ti...

Page 11: ...icians to do calibration of load cells The load cell calibration is very simple and fast with just one step 1kg weight calibration step Warning Misoperation in calibration could bring about serious co...

Page 12: ...message Click OpenFile to select update to update e g file RCC9A_V2 01 for main program update or Database_V2 02 for database update Then click Connect the PC displays the following message Click Pro...

Page 13: ...displayed in the left beneath part of each operation interface Upon 480 operations having been made the machine displays the following message and users can continue using the equipment meanwhile mak...

Page 14: ...service manual 2 Change compressor 3 Fasten pressure sensor connection to PCA or change the pressure sensor No change in recovery volume 1 No refrigerant in A C 2 Support screw of gas cylinder load c...

Page 15: ...s not well sealed Remove solenoid 8 from valve base clean the solenoid valve and valve base During vacuum there is suction in old oil bottle The contact between solenoid valve 2 and valve base is not...