This window can by opened in the Service program:
Menu -> Settings -> Setup kV Control…
Dosimetric Calibration Venus, CMP, RFX generators
, detailed information please find in the
enclosed documentation of the generator.
4.11
Exposure automatic
If the PROS 500 E is operated with an exposure automatic the functionality can be checked like this:
Place a Phantom or any other weakening object in the radiation way. Choose a measuring chamber
and expose. If this happens properly the measured value will be displayed. If something is not running
properly an Error message will be shown. Repeat this procedure for every measuring chamber.
•
The quickest exposure time of the Generator is 2ms.