PRS 500 F
Instructions for use
5067-0-8002
PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany
26 von 41
Detailed information please find in the enclosed User Manual of the X-ray tube.
4.5
Operation X-ray generator
Detailed information please find in the enclosed User Manual of the generator.
4.6
Operation Bucky, Grid entity
Detailed information please find in the enclosed User Manual.
4.7
Operation verticalX-ray system image receptor stand PROVERT
Detailed information please find in the enclosed User Manual PROVERT.
4.8
Operation RAPIXX system
Detailed information please find in the enclosed User Manual.
4.9
Operation Software
Detailed information please find in the enclosed installation- and User Manual CONAXX 2.
4.10
Function of the PRS 500 F
4.10.1
Switching On/Off the PRS 500 F
Switching on the PRS 500 F happens via the control panel of the Generator. The Generator supplies
every system component with power.
On the Generator and the control panel will run a self-test when switching them on. After the self-test
was successful the parameters will be displayed which can be saved under Organ-number #0.
When an Error gets displayed please see the manual generator.
Switch on the X-ray generator
button POW1
Switch of the X-ray generator
button POW2
ATTENTION!
Specially for the TOUCH option: The system is only switched off when the
touch display is completely shut down. A system restart is only possible after
the touch display has been shut down completely. Otherwise the touch
display will not start.
4.10.2
Dosimetric Calibration (only for PROVARIO HF)
The Generator is running a self –test, at switching on and changing the tube in a two tube operation,
on the chosen area dose measuring chamber.
On this test every Vacutec area dose measuring chamber emits a determinated
number of test pulses, which are getting detected by the Generator, the Generator compares it with a
deposited test value. At a dev/- 2% the error E018 „DAP-System“ gets displayed on the control
panel which indicates a decalibrated measuring chamber.
The test value of the area dose measuring chamber, which can be found on the test protocol, has to be
applied to the parameter 846 of the Generator or via the Service program into the field “DAP test count
value x (120-300)” (1).
The current measured test value is displayed in the field “DAP measured value x”(2) after switching on
the Generator. For this the function “read from gen”(3) has to be executed. Modest differences in the
measures test value can be caused by air pressure fluctuations or wrong installed measure chambers.