RAPIXX DR System Installation & User Manual
5420-0-0004
14 of 20
CONAXX 2 supports two different types of licensing:
Type of licensing
Properties
Detector-based
This type links the license with the detector.
Computer-based
This type links the license with the computer.
In case of having no question to choose the type of licensing, the detector supports computer-based
licensing only.
Note:
The module “Advanced image processing (AIP)” is always linked to the computer. It is not possible to use
it on other computers, even if the detector-based licensing is used.
There are two ways to send the hardware key:
a)
Online with the licensing page of the dealer backroom or
b)
by e-mail
For the online licensing, please login to the dealer backroom and navigate to the page “L
ICENSING
”. Select
the desired license and upload the hardware key with the function “S
ET
”. Download the license afterwards
with “G
ET
”.
For the licensing by e-mail, send the hardware key to
. You will then get the valid
license key by e-mail.
This license file can be imported with the function “Import license…” in the section “S
YSTEM
>
R
EGISTRATION
” of the CONAXX 2 configuration.
Now restart CONAXX 2. CONAXX 2 is now ready to acquire images with the detector. The proceeding
with the DR system is described in the document “CONAXX 2 User manual”.
Note:
For detailed information on CONAXX 2 licensing please refer to the document "CONAXX 2 User
Manual".
6.4
Setup dose indicator for detector
For each detector the document “RAPIXX calibration values” is included. Follow the instructions in the
document to setup the dose indicator.