RAPIXX DR System Installation & User Manual 

5420-0-0004 

 

 

 

10 of 20 

 

4

 

Installation 

4.1

 

Check list 

If PROTEC GmbH & Co. KG already prepared the installation, only the following sub-chapter from this 
chapter must be considered: 
 

 

4.2 Connection of the components 

Then proceed with chapter 

5 Calibration of the DR System 

on page 12. 

4.2

 

Connection of the components 

Connect the components as described in the following diagrams. 
 

WiFi

PC

CONAXX

Hand switch

X-ray generator

RAPIXX 2530V1V
RAPIXX 2530V2V

 

4.3

 

Setup the network 

During the installation of CONAXX 2 the “

E

B

US 

D

RIVER 

I

NSTALLATION 

T

OOL

” appears. It is also possible to start 

this tool via the start menu of Windows (A

LL PROGRAMS

\CONAXX2\DR-P

ANEL

\P

LEORA 

D

RIVER

T

OOL

). 

 
After starting the tool all detected network interface cards and wireless adapters are listed. Select the 
device with which the detector connection is to be established and select the entry “I

NSTALL 

U

NIVERSAL E

-

BUS 

D

RIVER

” in the column “A

CTION

”. Then press “I

NSTALL

”. Confirm all security messages of Windows and 

cancel

 all restart requests of Windows. After the successful installation close the driver tool. 

 
Go into “C

ONTROL PANEL

”, open „N

ETWORK AND SHARING CENTER

” and further “C

HANGE ADAPTER SETTINGS

”.  

 
Select the device with which the detector connection is to be established with the right mouse button 
and open the “P

ROPERTIES

”. Then select “I

NTERNET PROTOCOL 

V

ERSION 

4

 

/TCP/IP)” from the list.  By clicking 

“P

ROPERTIES

” you will get to a window, in which you can type in the IP address manually. 

Please enter there the IP address 

192.168.2.11

 and 

255.255.255.0

 for the subnet mask. 

 
The setup of the network is completed now. 
 

 

Summary of Contents for 4486-9-1000L

Page 1: ...hone 49 0 7062 92 55 0 E mail protec protec med com Version 2 0 Effective from 2020 06 30 Subject to alterations RAPIXX 2530V1V Wifi Set RAPIXX 2530V2V Wifi Set Digital X ray detector system Model ID...

Page 2: ...stenfeld Comments and questions about the documentation please contact PROTEC GmbH Co KG In den Dorfwiesen 14 71720 Oberstenfeld Germany Tel 49 0 7062 92 55 0 Fax 49 0 7062 92 55 60 E Mail protec prot...

Page 3: ...stallation 10 4 1 Check list 10 4 2 Connection of the components 10 4 3 Setup the network 10 4 4 W LAN connection to the Detector 11 4 5 Copying detector specific data 11 4 6 Installation of CONAXX 2...

Page 4: ...2 Safety related controls 18 9 3 3 Maintenance 18 10 Liability 18 11 Guarantee Fehler Textmarke nicht definiert 12 Disposal 19 13 Electromagnetic compatibility EMC according to EN 60601 1 2 19 13 1 Ge...

Page 5: ...pment as of the date of manufacture Revisions to the equipment subsequent to the date of manufacture will be addressed in service notes distributed by PROTEC GmbH Co KG Document Version Revision Date...

Page 6: ...persons responsible for the application must have the specific knowledge according to legal requirements and regulations and must establish safe exposure procedures for these kind of systems Those pe...

Page 7: ...D CONAXX 2 CD 1 3 System requirements For an optimal use of the DR detector it is necessary that the hardware fulfils the system requirements described in the document System requirements 1 4 Environm...

Page 8: ...ray detector systems of the RAPIXX series are only intended for the usage by professional users which are specially trained on the operation of diagnostic X ray units according to the correspondent n...

Page 9: ...lection of the used anti scatter grid 3 2 Notes for installing a new digital X ray system The generator cut off dose should be in the range between 3 0 and 3 3 Gy 3 3 Notes for upgrading an existing C...

Page 10: ...After starting the tool all detected network interface cards and wireless adapters are listed Select the device with which the detector connection is to be established and select the entry INSTALL UN...

Page 11: ...tically connect to the detector e g after battery changing 4 5 Copying detector specific data Install from the Detector CD the calibration data Normally the installation starts automatically If not pl...

Page 12: ...he window Click on OPEN LINK in the ACQUISITION menu Start the Gain calibration by clicking the button GAIN CALIBRATION in the ACQUISITION menu The detector now acquires 8 images without x ray radiati...

Page 13: ...ESULTS which is delivered with the detector The value is placed in the section SENSITIVITY in the field MAX COUNTS Depending on how the detector is built in it might be necessary to rotate the acquire...

Page 14: ...ing please login to the dealer backroom and navigate to the page LICENSING Select the desired license and upload the hardware key with the function SET Download the license afterwards with GET For the...

Page 15: ...close the toolbox If no error message appears at CONAXX 2 startup the connection is established correctly and the system is ready for use If no connection can be establish proceed as described in 7 3...

Page 16: ...during operation with the modality Should an error appear during the daily operation with the modality e g communication error power blackout the error message will show hints to solve the problem Is...

Page 17: ...e used such as Sani Cloth Plus a hospital grade EPA registered low to intermediate level product for hard non porous surfaces and equipment Use disinfectants in accordance with the manufacturer s inst...

Page 18: ...must be completed every 12 months by the PROTEC service department or a PROTEC authorized service provider in order to ensure the reliable operation of the unit As the manufacturer PROTEC is responsi...

Page 19: ...nd prevent any potential negative effects on human health and the environment which could otherwise arise from inappropriate waste handling In cases of any doubt contact PROTEC 13 Electromagnetic comp...

Page 20: ...according to EN 60601 1 Type B Electrostatic sensitive device Large focus Small focus System CONAXX 2 Modality CONAXX 2 X ray generator Electrical product Do not dispose in the household waste 14 2 A...

Reviews: