RAPIXX DR System Installation & User Manual
5420-0-0004
10 of 20
4
Installation
4.1
Check list
If PROTEC GmbH & Co. KG already prepared the installation, only the following sub-chapter from this
chapter must be considered:
4.2 Connection of the components
Then proceed with chapter
5 Calibration of the DR System
on page 12.
4.2
Connection of the components
Connect the components as described in the following diagrams.
WiFi
PC
CONAXX
Hand switch
X-ray generator
RAPIXX 2530V1V
RAPIXX 2530V2V
4.3
Setup the network
During the installation of CONAXX 2 the “
E
B
US
D
RIVER
I
NSTALLATION
T
OOL
” appears. It is also possible to start
this tool via the start menu of Windows (A
LL PROGRAMS
\CONAXX2\DR-P
ANEL
\P
LEORA
D
RIVER
T
OOL
).
After starting the tool all detected network interface cards and wireless adapters are listed. Select the
device with which the detector connection is to be established and select the entry “I
NSTALL
U
NIVERSAL E
-
BUS
D
RIVER
” in the column “A
CTION
”. Then press “I
NSTALL
”. Confirm all security messages of Windows and
cancel
all restart requests of Windows. After the successful installation close the driver tool.
Go into “C
ONTROL PANEL
”, open „N
ETWORK AND SHARING CENTER
” and further “C
HANGE ADAPTER SETTINGS
”.
Select the device with which the detector connection is to be established with the right mouse button
and open the “P
ROPERTIES
”. Then select “I
NTERNET PROTOCOL
V
ERSION
4
/TCP/IP)” from the list. By clicking
“P
ROPERTIES
” you will get to a window, in which you can type in the IP address manually.
Please enter there the IP address
192.168.2.11
and
255.255.255.0
for the subnet mask.
The setup of the network is completed now.