PRS-812 Resistance Meter
TABLE NOTES:
#1 AUTOMATIC:
RESISTANCE RANGES IN AUTO 0.1
TO <2.0E+12
as follows
@
<10V
: 0.1 TO <1.0E+04
(0.1
- < 10K
)
@
10V:
1.0E+04 TO <1.0E+06
(10K
- < 1M
)
@
100V:
1.0E+06 TO <2.0E+12
(1M
- <200T
)
#2 MANUAL:
RESISTANCE RANGES IN MAN 0.1
TO <2.0E+12
as follows
@
<10V
: 0.1 TO <1.0E+05
(0.1
- <100K
)
@
10V:
1.0E+03 TO <1.0E+09
(1K
- <1G
)
@
100V:
2.0E+05 TO <2.0E+12
(200K
-< 200T
)
#3 AUTO-MANUAL:
(Same as AUTOMATIC)
RESISTANCE RANGES IN AUTO-MANUAL 0.1
TO <2.0E+12
as follows
@
<10V
: 0.1 TO <1.0E+04
(0.1
- < 10K
)
@
10V:
1.0E+04 TO <1.0E+06
(10K
- < 1M
)
@
100V:
1.0E+06 TO <2.0E+12
(1M
- <200T
)
NOTE:
It is capable of measuring up to 2.0x10
14
ohms with tolerance depending on procedures and
conditions.
Resistance
2 Triangular button Arrows: UP
and DOWN
. Select Resistance
Range Select.
Range in single decades in Manual and Automatic/Manual modes
.
Test Volts
Manual Selection of <10, 10 or 100 volts in Manual Mode
Record/
Recall
Turns Memory Register ON if OFF. Provides access to all data in Memory
Register. Calculates and Displays Minimum, Maximum and Average of data
stored in Memory Register.
Clear
Erases data in Memory Register; if in HOLD mode, discards most recent Held
Value
ON/OFF
Power-up, perform functional & Battery tests, Power-down if ON.
Batt. Test
Displays
GOOD
on LCD if acceptable voltage or
Lo
if unacceptable
Reset
Enters (saves) data into Memory Register, Clears HOLD and Display.
Test
Begins
measurement sequence.
Battery Buss
ON/OFF Switch isolates batteries from instrument circuits for storage &
Cut Off
transport
Rev: PRS-812 – 4-2009
Format modification: 05-24-2005