PRS-812 Resistance Meter
7
In
AUTO-MANUAL,
the operator manually selects the initial resistance
range in decade increments. Once set, the PRS-801
starts
the
measurement process from the preset resistance decade, rather than re-
zeroing itself for each measurement. This feature saves measurement
cycle time and extends battery life. In this mode the instrument
automatically controls test voltage, range and measurement electrification
period.
[2] RESISTANCE
RANGE SELECT.
Two Triangular Arrow Buttons UP (
) and DOWN (
) select Resistance
Measurement Range in single orders of magnitude while in MANUAL or
AUTOMATIC/MANUAL modes
[3] TEST VOLTS
Manual Selection of <10, 10 or 100 volts in
MANUAL,
selects
initial
test
voltage.
MANUAL
Test Voltage & Resistance Limits are as follows:
@
<10V
: 0.1 TO <1.0E+04
(0.1
-< 10K
)
@
10V:
1.0E+02 TO <1.0E+09
(100
-< 1G
)
@
100V:
1.0E+05 TO <2.0E+14
(100K
-< 2T
)
NOTE:
Optimal
minimum measurement in
MANUAL
using 100 volts is 2.0E+05.
[4] RECORD/
RECALL
If
REC
is not displayed
in the lower left corner of the LCD, pressing
RECORD/RECALL
once turns Memory Register ON [
REC
will then be
displayed in LCD]. If
REC
is ON and measurement data is stored in
memory, pressing
RECORD/RECALL
once will provide access to the
memory register. Pressing
RECORD/RECALL
successively will
calculate and sequentially display Minimum [MIN], Maximum [MAX] and
Average [AVG] of data stored in the Memory Register. If memory is ON
and register contains stored data pressing
RECALL
will provide the
following information:
First
Press
RECALL
:
Provides access to data in
Memory
NOTE:
MEM
XX Flashes in LCD indicating number of data points in the
register and the last measurement is displayed. Pressing DOWN
(
)
and
UP
(
)
displays other data points and their respective position in the
register is shown in the
MEM
XX section of the LCD.
Second
Press
RECALL
: Displays
MIN
Data Point in Memory
Third
Press
RECALL
: Displays
MAX
Data Point in Memory
Note:
When displayed,
OL
[Over Level] indicates a
measurement
greater than
(>)
2.0x10
14
ohms, which is beyond
the measurement capability of the PRS-812.
Fourth Press RECALL
: Displays
AVG
of all Memory Data Points